HPG 13150 400 V (A)
Troubleshooting and Customer Support
6-7
HPG™ 13150 400 V RF Power Generator
Prerelease version
Error Descriptions and Resolutions.
lists the error messages that may occur on the HPG generator,
describes each error message, and explains how to resolve the error. If the
information in this table does not resolve the problem, contact Customer
Support (see
“Customer Support” on page 6-21
).
Table 6-1. Error description and resolution
Error Message
Description and Resolution
E01 Interlock Loop
Open
Open interlock loop.
Causes: Switch opened, interlock loop not connected.
Solution: Close interlock loop by making the proper
connections to satisfy the interlock. See pins 10 and 23 in
E02 Water Flow Too Low
Cooling water flow below limit.
• HV and filament are switched off.
• HPG will reinitialize after error condition is resolved.
Cause: Water flow detector reports insufficient water flow.
Solution: Increase cooling water flow.
E04 Mains Failure
Softstart reported failure.
• HV is switched off.
• HPG will reinitialize after error condition has gone.
Cause: Mains interrupted, phase missing.
Solution:
1. Switch HPG generator off and check mains.
The
ON/OFF
power switch does not completely
disconnect the
Mains
. You must have an external switch
installed to completely disconnect the
Mains
.
2. Turn unit back on.
3. If problem persists, contact customer service.
Summary of Contents for 13150 400 V
Page 2: ......
Page 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Page 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...