Maintenance instructions
Replacing the external batteries
14. Secure the cover by the screw (
Fig. 64
/1).
Fig. 64
Detail view of trolley; screw
15. Insert the connection socket of the DS battery cable for the external bat-
teries in the connector on the rear of the Savina 300.
Fitting of the external batteries is complete.
16. Perform the „function test“ and „electrical safety test“ as per the test
instructions.
17. Fully charge the external batteries (for time taken see instructions for
use).
CAUTION
Batteries not fully charged! If the batteries are not fully charged, the running
time of the device in the event of a mains power failure may be reduced
considerably.
–
Inform the user that the rechargeable batteries still need to be
recharged.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...