Maintenance instructions
Replacing the microfilter
3. Withdraw the microfilter (
Fig. 30
/2) by the tab (
Fig. 30
/1) out of its mount
and dispose of it in the domestic waste bin.
Fig. 30
Savina 300 rear; microfilter
Removal of the microfilter is complete.
Fitting
1. Insert the new microfilter (
Fig. 30
/2) into its mount as far as it will go.
2. Close the tab (
Fig. 30
/1) against the microfilter.
3. Slot the pegs on the underside of the filter cover into the recesses in the
rear panel, swivel the filter cover upwards and with the catches (
Fig. 29
/1)
press the filter cover into the rear panel until it locks in place.
4. Lay the power cable (
Fig. 29
/2) around the filter cover guide, as shown in
the illustration.
Fitting of the microfilter is complete.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...