41 / 50
Function and condition test
Function tests
Entry
Measured value (external O2 meter)
[________vol% 02]
Action
• Set specified O2 value to 21 % by volume O2.
4.2.8
Paw pressure regulation (VC-CMV/VC-AC)
Prerequisites
The Savina 300 is connected to the mains power supply and to the „O2“ com-
pressed gas supply and switched on. The Savina 300 is set to „VC-CMV/VC-
AC“ mode.
Action
• Prepare the following test setup:
Test set-up
Fig. 14
Test setup; Paw pressure regulation
Action
• Press the „Ventilation settings“ softkey and set the following parameters:
– Ti : 2.0 s
– f: 6/min
– PEEP: 5 mbar
Item
Designation
1
Pressure gauge
2
Test lung
3
Y-piece, adult, Luer-Lock
4
Adult ventilation tubing
5
O2 analyzer
6
O2 adapter
NOTE
The display message „VTi low“ on a yellow background may be triggered
during this test. It does not affect the result of measurement.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...