39 / 50
Function and condition test
Function tests
Action
• Press the „Alarms“ softkey and set the following parameters:
– MV high: 41 L/min
– MV low: 0.5 L/min
– Paw: 100 mbar
– VTi: 4.00 L
– f: 120/min
– Tapn: 60 s
Action
• Press the „Sensors/Parameters“ softkey and set the following parameters:
– O2 monitoring: On
– Flow monitoring: On
Action
• Press the „System configuration“ softkey.
• Press the „Ventilation“ softkey and set the following parameters:
– Pressure limitation: Off
– Plateau: On
– LPO: Off
• Start ventilation with the Savina 300.
O2 calibration
Action
• Select the „Sensoren/Parameter” menu.
• Press the O2 calibration „Start“ softkey and confirm with the rotary knob.
Test
The message „Disconnect patient“ is displayed.
Action
• Disconnect the ventilation tube from the inspiratory socket.
Test
The display shows „O2 calibration in progress“ and then „Reconnect patient“.
Action
• Reconnect the ventilation tube to the inspiratory socket.
Test
The following message is displayed: „O2 calibration OK“
Entry
[________OK]
VT flow control and VT measurement
Action
• While the test lung is being ventilated, push the Spirolog sensor to the left-
hand end position and back to the right-hand end position in order to trig-
ger a flow calibration.
• After the display message „Flow calibration OK“ appears, ventilate the test
lung for at least another three cycles.
• Read out the „VTe“ measured value and compare it with the specified toler-
ance range.
VT=500 mL (expiratory valve 1)
The displayed value is in the range from
440
mL to
540
mL.
Entry
[________mL]
Action
• After the tidal volume measurement (VT) perform a flow calibration.
NOTE
Perform the „VT metering and VT measurement“ test with
all
the expiratory
valves listed in the „Device configuration“ section under „Serial numbers“.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...