Electrical safety
Electrical safety to EN 62353
20 / 50
6DYLQD
5HYLVLRQ
3.1.6
Leakage current, mains on applied part with airway temperature sen-
sor AWT 01, if installed
Prerequisites
The tester is switched on.
Action
• Prepare the following test setup:
Test set-up
Fig. 3
Leakage current of airway temperature sensor
Action
• (applied part) Connect the airway temperature sensor on one end to the
device under test and on the other end by a test clip with measuring lead
and the tester, configurable socket „A“ for applied parts (paying attention to
the configuration!).
• Follow the instructions on the tester.
NOTE
In the following test the leakage current is measured at the airway tempera-
ture sensor AWT01. The expected value is very low (the typical measured
value is 1.5 μA to 2 μA).
Item
Designation
1
Tester
2
Device under test
3
Connection socket of device under test
4
Airway temperature sensor AWT 01
5
Test clip with measuring lead
6
Configurable sockets for applied part
L
Conductor
N
Neutral conductor
PE
Protective earth conductor
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...