Electrical safety
Electrical safety according to IEC 60601-1
24 / 50
6DYLQD
5HYLVLRQ
3.2
Electrical safety according to IEC 60601-1
3.2.1
Visual check
Prerequisites
Savina 300 is switched off and not connected to the mains power supply.
Action
• Check the following items for damage:
– Power supply cord of the device
– Power switch
– Power fuse link for mains power supply
– Fuse link for internal battery
Test
The items mentioned above are undamaged. The fitted power fuse-links
match the values specified on the labels.
Result
Visual check completed.
[________OK]
NOTE
The medical product to be tested conforms to the requirements of protection
class I, type B with an airway temperature sensor type BF connected.
NOTE
An optional multiple socket-outlet, if any, must be included in the individual
tests (medical electrical system).
NOTE
With devices that are connected to other devices by means of a data cable,
this connection must be disconnected prior to performing the electrical safe-
ty check, in order to avoid incorrect measurements.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...