N
o
.02
08
_00
00
00
253
3_
Pu
bl
ic
ati
on
6DYLQD
Function and condition test
Function tests
36 / 50
6DYLQD
5HYLVLRQ
4.2.4
Power supply unit
Electrical supply display
Prerequisites
The Savina 300 is connected to the mains power supply.
Action
• Set the Savina 300 to „Standby“ mode.
Test
The LEDs on the front indicate the following operating states:
– „Mains power supply“ LED lit green; „External battery“ LED out
– „Internal battery“ LED lit „green“ when internal rechargeable batteries fully
charged or lit „yellow“ when the internal rechargeable batteries are charg-
ing.
Entry
[________OK]
Internal supply voltage (internal battery)
Prerequisites
Savina 300 is in „Standby“ mode.
Action
• Cut the mains power supply to the Savina 300.
Test
The acoustic alarm sounds. The message "Internal battery activated" is dis-
played on a yellow background. The LEDs on the front of the device indicate
the following operating states:
– „Mains power supply“ LED out; „External battery“ LED out
– „Interne battery“ LED lit „green“ or lit „yellow“
Action
• Press the „Start/Standby“ key and switch the Savina 300 to „VC-CMV/VC-
AC“ mode.
Entry
The blower runs up to speed. Ignore the error messages and alarms during
this test.
[________OK]
Action
• Connect the Savina 300 to the mains power supply.
Power failure alarm
Action
• Remove the fuse for the internal rechargeable battery. The fuse is located
on the rear panel of the unit.
Test
The message "
Internal
battery
is missing
" is displayed on a red background.
Action
• Press the „Alarm Reset“ key.
• Cut the mains power supply to the Savina 300.
Test
All LEDs on the front are out. The acoustic power failure warning sounds.
Action
• Switch off Savina 300.
Test
The acoustic alarm stops.
Action
• Re-insert the fuse for the internal rechargeable battery.
• Restore the mains power supply to the Savina 300.
• Switch on the Savina 300.
• Set the Savina 300 to „Standby“ mode.
NOTE
For the following tests do not connect any external battery or DC supply to
the Savina 300.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...