Electrical safety
Electrical safety according to IEC 60601-1
3.2.2
Protective conductor resistance
Test set-up
Fig. 5
Test setup for protective earth conductor resistance
Action
• Prepare the test setup.
• Switch on the tester and the device under test.
• Configure the tester appropriately, and follow the instructions on the tester.
• Using the tip of the test probe, scan the following points on the device
under test, move the power supply cord section-wise while doing so:
– Power supply unit earth pin
– Screws on the housing
– Oxygen connection
– Optional rails (on side of Savina)
– LPO option (connector)
Test
The protective conductor resistance must not exceed
0.2
Ohm (including
mains power cable) in each case.
Entry
Enter the highest measured value of the protective conductor resistance.
[________Ohm]
NOTE
The protective earth conductor resistance is measured with the power cable
connected.
Item Designation
1
Tester, e.g. SECUTEST
2
Device under test
3
Test probe with tip
L
Conductor
N
Neutral conductor
PE
(Protective Earth) conductor
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...