Electrical safety
Electrical safety to EN 62353
3.1.8
Optional multiple socket strip on trolley
Power fuse links
Action
• Compare the power fuse links with rated values on the multiple socket
strip.
Test
The power fuse-links match the specifications on the multiple socket strip.
Result
[________OK]
Protective conductor resistance
Action
• Check the protective conductor contacts of all sockets on the multiple
socket strip with tester probe.
Test
The value of the protective earth conductor resistance must not exceed
0.3
ohms in each case.
Entry
Document the highest protective conductor resistance measurement in the
„Test Report“.
[________Ohm]
Equivalent device leakage current
Test
The initial value must not exceed
100
μA.
Entry
Initial value
[________μA]
Test
The recurrent test value must not exceed
100
μA.
Entry
Recurrent test
[________μA]
NOTE
The equivalent device leakage current test applies to the multiple-socket
outlet only!
NOTE
Always enter the initial value in a new „Test Report“ document.
Summary of Contents for Savina 300
Page 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Page 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Page 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Page 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Page 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Page 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Page 110: ... This page has been intentionally left blank ...
Page 164: ... This page has been intentionally left blank ...
Page 165: ... This page has been intentionally left blank ...