Savina 300
Maintenance instructions
Replacing the O2 sensors
4. Dispose of the O2 sensors according to local waste disposal regulations.
Removal of the O2 sensors is complete.
Fitting
1. Push the O2 sensors (with the arrows on the oxygen cells pointing
upwards, as shown in the following illustration) into their mounts until you
feel a resistance.
Fig. 38
Device front; fitting the O2 sensors
2. Swivel the inspiratory port (
Fig. 39
/1) down.
3. Use the screw (
Fig. 39
/2) to secure the cover plate (
Fig. 39
/3) on the
Savina 300.
Fig. 39
Device front; cover plate
Fitting of the O2 sensors is complete.