MQSA Quality Control Tests | 65
Running the Quality Control Tests for Medical Physicists
The Quality Control tests listed in this section must be performed by
the medical physicist to ensure that the Dome display provides a high
level of mammographic image quality.
These tests also establish the basis of the equipment evaluation
that must be performed following the installation of a Dome display,
the repair or replacement of a major component of the system,
the relocation of a Dome display to another viewing location, or
the substantial change of illumination in an existing viewing
environment.
You must conduct the medical physicist series of quality control tests
at least annually. Use the test configuration dialog to set up test
reminders, or to set up reminders to conduct the test series more
frequently, quarterly or biannually.