90 | Dome CXtra
• Conformance. Presents values of Just Noticeable Differences (JND)
per luminance interval and the mean value for JND/luminance
interval. These values include the Luminance Uniformity Metric
(LUM - DICOM metric) value and the FIT line expression used to
quantify the degree of conformance.
In general, the closer the JND values are to the mean line, the better
the DICOM conformance. In addition, the closer the calculated LUM
metric is to 0 (zero), the better the DICOM conformance.