48 | Dome CXtra
MQSA Quality Control Tests
Regular inspection and testing of diagnostic imaging displays
designated for primary image diagnosis in mammography ensure
display calibration and quality control. Only the Dome E5 and
Dome S10 displays can be used for image diagnosis as regulated by
the Mammography Quality Standards Act, or MQSA.
The Dome CXtra software helps you set up test reminders for
consistent quality control activity and provides a way for you to
record the results of the quality control tests electronically.
On the Dome E5 display, the MQSA Test Wizard is accessible only via
the CXtra Services menu or the Windows Start menu.
With the Dome S10 display, you can access the wizard via the MQSA
button as well as by the CXtra Services menu or Windows Start menu.
A mini-USB cable connecting the display to the computer system is
required for button functionality.
Scope
The quality control tests are appropriately applied to the displays
when they are specifically used for the clinical review of images
acquired by full-field digital mammography systems, also called
FFDM systems.
Materials/Equipment
The following items are required for the quality control tests.
• Dome E5 or Dome S10 display system with graphics controller and
driver installed
• Computer workstation running Microsoft Windows XP
Service Pack 3 or later
• Dome CXtra software