RightLight Service | 27
External Sensor tab
Specify the external sensor tests that you want to complete manually.
For each manual test that you select, you can specify a threshold
value that defines a test failure. When a failure occurs, the CXtra icon
is set to a warning state. In addition, if the CXtra-enabled workstation
is connected to Dome Dashboard, an alert is also sent to the
Dashboard app or SNMP console. Selections made on this tab
determine operating parameters for the Conformance Wizard, which
guides the manual test.
To set alert parameters
1 Select either the RightLight White Level test or the DICOM LUM test
or both, for external sensor conformance testing.
• RightLight White Level test. Check the selection box, then
enter a value for Tolerance ±. The RightLight White Level test is
classified a failure if the measured white level is outside the range
defined by this percent, applied to the Target Luminance defined
on the Backlight tab. Enter a value ranging from 0 (zero) to 25
(default =10.00).
• DICOM LUM test. Check the selection box, then enter a value for
Maximum allowed. The DICOM LUM test is classified a failure if it
produces a value greater than the maximum allowed. Enter a
value ranging from 0 (zero) to 2 (default=1.00). A value of 0 (zero)
indicates perfect conformance with the DICOM Grayscale
Standard Display Function.