![Dome CXtra User Manual Download Page 6](http://html.mh-extra.com/html/dome/cxtra/cxtra_user-manual_2517705006.webp)
2 | Dome CXtra
Expansion of Manual Conformance Testing
With the addition of the front sensor on some models of Dome
displays, the Manual Conformance functionality has been bifurcated
into the External Sensor Conformance testing (using the photometer
device) and the Front Sensor Conformance testing.
External Sensor Conformance replaces Manual Conformance on the
CXtra Services menu and the Conformance wizard. All operational
procedures are unchanged.
Front Sensor Conformance has been added to the menu. With this
functionality you can perform auto and manual conformance testing.
NOTE: Applies to Dome S3, Dome S3c, and Dome S10 displays only.
White Level Maximum Limit Exceedance
If you have a requirement to exceed the maximum limit of the preset
White Level target range, you can now do so but your display
warranty becomes void. Contact NDSsi Customer Care for more
information.
European Reference Guidelines
The EUREF guidelines, organized by the European Reference
Organization for Quality Assured Breast Screening and Diagnostic
Services, promotes mammography care of high quality in European
Union member states. Users of Dome displays qualified for AAPM
TG18 can perform the White Level Calibration procedure to meet this
standard. See “White Level Calibration Service” on page 44.
Modification of Color Calibration Setting
Only an administrator, not privileged users, can modify the color
calibration setting accessible on the RightLight Panel Configuration
dialog, Color tab. (See page 22.)
NOTE: Applies only to Dome displays using the RightLight
functionality.