MQSA Quality Control Tests | 51
To configure the MQSA test reminders
1 Open the CXtra Services menu and select MQSA Test Configuration.
2 Optional: Check the selection box to enable Expert Mode. This
mode allows radiological technicians and medical physicists
already familiar with evaluation of the SMPTE test pattern to
quicken visual inspection of the display screen.
3 Check the selection box to enable the functionality for MQSA test
reminders.
NOTE: Reminders are issued only for the test(s) that you enable on
the dialog.
4 Enter a time period for test postponement. Input a number from
1 to 59 for the number of minutes, or a number from 1 to 24 for
the number of hours.
NOTE: When using the Dome S10 display, you can click the Ignore
Test button on the wizard to temporarily reschedule a reminder.
The reminder snoozes for the length of time specified. Notice that
the button face toggles to Reset Ignore Test. Click the button again
to restart the countdown of the test postponement time.
5 Set the daily test reminder. Use the pull-down arrow to select a
time for any or all weekdays. Click the corresponding selection
box(es) to enable the daily test for the day(s). On the Dome S10
display, configure the warning light indicator. See “
To configure the
warning lights for test reminders
” below.
6 Set the monthly test reminder. Enter a day or date. When you select
a day, you also need to choose an ordinal value. Click the selection
box to enable the test. Configure the warning light indicator, if
using a Dome S10 display. See “
To configure the warning lights for
test reminders
” below.
7 Set the quality control test reminder for medical physicist. Select
the frequency of the tests, either quarterly, biannually, or annually.
Click the selection box to enable the test. Set the starting month for
the tests, then enter a day or date. When you select a day, you also
need to select an ordinal value. Configure the warning light
indicator, if using a Dome S10 display. See “
To configure the
warning lights for test reminders
” below.
8 Click OK to finish.