![Dome CXtra User Manual Download Page 157](http://html.mh-extra.com/html/dome/cxtra/cxtra_user-manual_2517705157.webp)
Enterprise Management Service | 153
Display Agent Traps
Systems running the Enterprise Management service for Dome CXtra
generate only the following traps:
•
trapStatusObserverConnectionFailed
•
trapStatusChanged
If the local Dome CXtra system status changes to Warning or Failure,
the display agent sends a
trapStatusChanged
trap. For example,
if an external application changes the calibration from DICOM
Calibration to Original Calibration, the DICal service changes the
Dome CXtra system status to Warning and sends the
trapStatusChanged
trap.
To have traps sent when the Dome CXtra system status changes to
Success, set
snmpSuccessTraps
to “true” in the
sysDOMEinfo
table.
See the following table of available traps.