28 | Dome CXtra
2 Select the number of luminance measurements you want
to make during each manual test.
Note: For the external sensor, the default number of measurements
is 15. When you select All, 256 measurements are taken on 8-bit
display systems; 1024 measurements are taken on 10-bit systems.
If you run the External Sensor Conformance Wizard with the selection
boxes unchecked for both the DICOM LUM test and the White Level
test, you are logging the data from a completed manual conformance
test without applying the criteria of the DICOM LUM and White Level
tests for Pass/Fail determination.
Running Conformance Tests
After you configure the RightLight service, the conformance tests that
you specified on the Front Sensor, Rear Sensor, and External Sensor
tabs are set to run. That is, per your selections, RightLight performs
White Level tests or DICOM LUM tests or both where auto and manual
alerts are set accordingly for front, rear, and external sensor
conformance testing.
ALERT
Front
Sensor
Rear
Sensor
External
Sensor
Auto
> White Level
> DICOM LUM
9
Auto-scheduled
> RightCheck White Level
> DICOM LUM
9
Manual
> White Level
> DICOM LUM
9
9