![Dome CXtra User Manual Download Page 34](http://html.mh-extra.com/html/dome/cxtra/cxtra_user-manual_2517705034.webp)
30 | Dome CXtra
Auto Conformance Testing via the Rear Sensor
Automatic conformance testing shows how well your display
conforms to the DICOM standard. RightLight automatically performs
the tests selected on the Rear Sensor tab every 5 minutes. See
page 25 to view or modify the test parameters.
If you have the Privilege service installed, you must log on as a
privileged user before you set up auto conformance testing. (See
page 138.)
Checking Conformance via the External Sensor
If you have configured RightLight properly and have an external
photometer installed, you can run a manual external sensor
conformance test.
If you have the Privilege service installed, you must log on as
a privileged user before you run a manual conformance test.
(See page 138.)
As with automatic conformance testing, the manual conformance
test shows how well your display conforms to the DICOM standard.
You are able to specify, however, the number of measurements to be
taken. See “External Sensor tab” on page 27 for configuration
information. See “Installing the Photometer” on page 10 for how to
install the photometer.
Before starting a manual conformance test, select Dome RightLight
DICOM Calibration on the services menu.
NOTE: If you run a conformance test with another response function
(such as Dome RightLight Exponential Calibration or Dome
RightLight Native Calibration), the system presents a warning
message. Continuing to run the test causes the system to show
the CXtra icon in the warning state.