4 • Troubleshooting
Doc. 031644-02 9/03
4-11
•
Incorrect reference wavelength
Select a reference wavelength where minimum absorbance occurs. Make sure
the reference wavelength is included in the wavelength range.
•
Wavelength range too narrow
Make sure the spectrum wavelength range includes both the reference
wavelength and reference bandwidth. Make sure the peak absorbance
wavelength is not included in the reference bandwidth.
•
Reference bandwidth too wide
Select a narrower reference bandwidth. The reference bandwidth should not
overlap the sample bandwidth.
4.12 Peaks Too Large or Small
•
Analog output-related problem
Check the recorder and integrator input voltage. The PDA-100 provides a
1 volt full-scale output.
Verify that the selected output range (AU full-scale) is appropriate.
•
System-level problem
Verify that the sample volume or concentration is appropriate.
Verify that the rise time and data collection rate are appropriate to capture the
narrowest peak. See Section 3.4 for more information.
Prepare fresh post-column reagent.
Verify that the reagent flow rate is correct.
•
Reference bandwidth too wide
Select a narrower reference bandwidth. The reference bandwidth should not
overlap the sample bandwidth.