Wavelength calibration
Reassigns new wavelengths to each pixel, based on the emission
spectrum from the deuterium lamp. Both the initial calibration and
previous calibration are stored in the calibration log and can be
retrieved at any time.
To calibrate the wavelength:
a. Press F8 to display the Commands dialog box.
b. Select
UV
(to display UV-related commands).
c. Select
UV_Calibration
and then
WavelengthCal
.
d. Click the
Execute
button.
e. When the calibration procedure is complete, select the
WavelengthCalResult
command, or click the
Wavelength
button on the PDA-100 Wellness control panel.