5 • Service
Doc. 031644-02 9/03
5-11
5.9
Replacing the Main Power Fuses
1. Turn off the PDA-100 power switch.
2. Disconnect the power cord from the PDA-100 rear panel.
3. The fuse holder is part of the main power receptacle (see Figure 5-4). Note the
recessed lock on each side of the fuse holder; now, using a small screwdriver
(or your fingernails), push one of the locks toward the center to release it. The
fuse holder pops out slightly when the lock is released. Repeat for the other
lock.
HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the PDA-100.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau arrière du PDA-100.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des PDA-100.
Figure 5-4. Main Power Fuse Holder
Fuse Holder
Fuse Holder
(Side View)
Fuses (2)
Insert
screwdriver
and twist to
release
(each side)
Locking
Spring
Locking
Spring
Main Power
Receptacle
Key