Doc. 031644-02 9/03
A-1
A • Specifications
A.1 Physical
A.2 Environmental
A.3 Electrical
Dimensions
17.5 cm H x 38.6 cm W x 49.0 cm D
(6.9 x 15.2 x 19.3 in)
6 cm (2.4 in) clearance required behind the detector
Weight
16 kg (35 lbs)
Decibel Level
<50 db (at “A WEIGHTING” setting) during routine operation
56 db (at “A WEIGHTING” setting) at power-up
Operating
Temperature
4 ºC to 40 ºC (40 ºF to 104 ºF)
Operating
Humidity
5% to 95% relative humidity, noncondensing
Main Power
85 Vac to 264 Vac, 47/63 Hz. The input power is 162 W,
maximum; the output power is 130 W, maximum. The
maximum input current is 4 amps.
The auto-sensing power supply requires no voltage or frequency
adjustment.
Fuses
Two 4.0 amp fast-blow IEC 127 fuses (P/N 954763)