PDA-100 Photodiode Array Detector
2-6
Doc. 031644-02 9/03
2.3.1
Flow Cell
The PDA-100 flow cell assembly (see Figure 2-4) is available in PEEK or
stainless steel. Two built-in heat exchangers, one on the cell inlet tubing
and one on the cell handle, help stabilize the temperature of the mobile
phase before it enters the cell.
Figure 2-4. Flow Cell Assembly
Flow Cell
Cell Path Length
Cell Volume
Heat Exchanger Volume
PEEK
(P/N 056346
)
10 mm
13
µ
L
20
µ
L
Stainless Steel
(P/N 056126)
10 mm
13
µ
L
9
µ
L
Do not touch the cell windows. If you touch a window, clean it with
denatured alcohol and a clean lens tissue.
Do not use the PEEK flow cell with normal phase or chlorinated
solvents; these solvents will damage the cell.
Do not use the stainless steel flow cell with high pH salts; the salts
can cause corrosion, contamination, and metal leaching.
Cell Inlet Tubing
Cell Outlet Tubing
Heat Exchangers
Flow Cell
Handle
Spring Clip