PDA-100 Photodiode Array Detector
3-14
Doc. 031644-02 9/03
The following screen is displayed.
14. Program the single-wavelength channels by selecting the
Wavelength
,
Bandwidth
,
Reference Wavelength,
and
Reference
Bandwidth
for each channel.
NOTE The settings selected for UV_VIS_1 through
UV_VIS_4 apply to both analog outputs and digital
outputs. (UV channel 1 = analog output 1, UV2 =
analog output 2, etc.). Specify the Recorder Range
(in mAU per volt) for each analog output that will
be collected.
NOTE When the Reference Wavelength is Off, the
Reference Bandwidth is not used and the sample
wavelength and sample bandwidth are used as
reference parameters.
15. After selecting all PGM file parameters, click
Finish
.