PDA-100 Photodiode Array Detector
3-8
Doc. 031644-02 9/03
PeakNet 6 and Chromeleon include several detector commands. These commands
are selected from the default control panel or the Commands dialog box (selected
from the
Control
menu or by pressing F8).
The table below lists several of the most frequently used detector commands. For
more information, refer to the online Help.
PDA-100 Command
Function
Connect
Indicates whether the PDA-100 is connected to the server
and the selected timebase.
UV Lamp On/Off
Turns the deuterium lamp on and off.
Vis Lamp On/Off
Turns the tungsten lamp on and off.
UVLampPowerOnSetting
Selects the power-up setting for the deuterium lamp.
VisLampPowerOnSetting
Selects the power-up setting for the tungsten lamp.
Autozero
Adjusts the output to zero.
Recorder_Range
Sets the full-scale recorder output range.
Offset_Level
Sets the offset applied to the recorder output.
Reference_Mode
Reports the reference mode.
RefWavelength
Sets the reference wavelength. If RefWavelength is
Off
,
the reference value is the first data point of the sample
wavelength and bandwidth after autozero is performed.
RefBandwidth
Sets the bandwidth of the reference wavelength, if one is
selected. If RefWavelength is
Off
, the RefBandwidth
setting has no effect.
Wavelength
Sets the sample wavelength (up to five sample
wavelengths can be programmed).
Bandwidth
Sets the bandwidth of the sample wavelength.
BunchWidth
Bunching (or averaging) the signals of adjacent
wavelengths reduces the size of the 3D data stored.
NegativeAbsorptionLevel
Adjusts the negative absorbance; this permits peaks to be
seen even when the eluent decreases absorbance.
WavelengthCal
Runs the wavelength calibration procedure.