2 • Description
Doc. 031644-02 9/03
2-9
Data Processing Card
The PDA-100 Moduleware resides on the Data Processing card. Four BNC
connectors on the card provide the analog (recorder) outputs. Select the output
scale, polarity, baseline offset, and filter rise time from PeakNet 6 or Chromeleon.
In addition, the Data Processing card provides the digital signal processing and
conversion to log ratios when absorbance is measured.