19.2 General Specifications – LEEP System 1000
®
Electrosurgical Generator
ENVIRONMENTAL CONDITIONS
(Usage)
ENVIRONMENTAL TEMPERATURE . . . Between 50 °F and 113 °F (10 °C and 45 °C)
RELATIVE HUMIDITY . . . . . . . . . . . . . . Between 30% and 75%
(Shipping and Storage)
ENVIRONMENTAL TEMPERATURE . . . Between -40 °F and 158 °F (-40 °C and 70 °C)
RELATIVE HUMIDITY . . . . . . . . . . . . . . Between 10% and 100%
PHYSICAL - Electrosurgical Generator
DIMENSIONS . . . . . . . . . . . . . . . . . . . . 13" wide x 14" deep x 7¼" high (330 mm x 356 mm x 184 mm)
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . . . 30 pounds (14 kg)
ELECTROSURGICAL OUTPUT
RF OUTPUT FREQUENCY . . . . . . . . . . 450 kHz
RF OUTPUT POWER:
Volts p-p Max Duty Crest
(open circuit) Cycle Factor
CUT 0-100 watts RMS* 830 --- 1.4
BLEND 0-100 watts RMS* 1200 60% 2.0
COAG 0-80 watts RMS* 3800 10% 5.5
*Stable to >800 ohms (calibration at 500 ohms)
RF ISOLATION . . . . . . . . . . . . . . . . . . . . Less than 150 milliamps at 200 ohms
CLASSIFICATION . . . . . . . . . . . . . . . . . I-Type BF
OUTPUT CIRCUIT . . . . . . . . . . . . . . . . . Floating output. Protected against the effects of the defibrillator.
WORKING MODE . . . . . . . . . . . . . . . . . Discontinuous maximum duty cycle: 10/30 sec.
COOLING . . . . . . . . . . . . . . . . . . . . . . . . Convection cooling without fan
CONTROL . . . . . . . . . . . . . . . . . . . . . . . Foot pedal operated (pneumatic) with audible signals
and mode lights
AUDIBLE SIGNALS AND LIGHTS FOR OPERATION AND ALARM:
MAIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . Green light
ALARM, PATIENT PLATE CONTINUITY . Low pitch intermittent audible alarm - red light
ALARM, OUTPUT POWER . . . . . . . . . . Higher pitch intermittent audible alarm
CUT AND BLEND MODES . . . . . . . . . . . Low pitch audible signal - yellow light
COAGULATION MODE . . . . . . . . . . . . . High pitch audible signal - blue light
NOTE: Specifications subject to change.
OUTPUT POWER AT 500 OHMS:
Pure Cut
100 W RMS (open circuit 830 Vp-p; crest factor 1.43)
Waveform:
sinusoidal at 450 kHz
Blend Cut
100 W RMS (open circuit 1200 Vp-p; crest factor 2)
Waveform:
sinusoidal at 450 kHz
Duty cycle:
60%
Coagulation
80 W RMS (open circuit 3800 Vp-p; crest factor 5.5)
Waveform:
sinusoidal at 450 kHz
Duty cycle:
10%
28
LEEP System 1000
®
Workstation • Model KH1000A • Operating Manual • English
(continued)
6
7
5
Summary of Contents for LEEP System 1000
Page 4: ...This page left blank intentionally 2 ...
Page 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Page 36: ...Leerseite 34 ...
Page 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Page 100: ...Page intentionnellement vierge 98 ...
Page 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Page 164: ...Página intencionalmente em branco 162 ...
Page 195: ......