2.5 Connecting the Smoke Evacuator and LEEP System 1000
®
Electrosurgical Generator
together and to the Wall Outlet
1.
Plug in the cord running from the Electrosurgical Generator to the receptacle at the back of the
Smoke Evacuator [refer to (A) in Figure 1].
2.
Make sure both On/Off Switches on the Electrosurgical Generator panel are in the OFF position.
Refer to and in Figure 2. Next, attach the power cord to the receptacle at the back of the Electrosurgical
Generator and then into a hospital-grade grounded wall outlet to obtain power [refer to (B) in Figure 1].
2.6 Installing the Foot Pedal Switch on the LEEP System 1000 Electrosurgical Generator
Connect the Foot Pedal Switch to the socket , shown in Figure 2, without activating the Foot Pedal, and
tighten the threaded plug. This is an air (pneumatic) operated control. There is no electric current, offering
maximum safety.
2.7 Installing the Electrodes and Dispersive Patient Plate
2.7.1 Placement of the Electrode
•
Connect the active Electrode Handpiece to the socket , shown in Figure 2, and tighten the
electrode of choice in the Handpiece.
2.7.2 Placement of the Patient Plate or Dispersive Electrode
When using an electrosurgical system, it is very important that all the current delivered to the
patient returns correctly to the Electrosurgical Generator via the Dispersive Patient Plate only.
•
Connect Dispersive Patient Plate to the socket . Refer to Figure 2.
•
The patient must be positioned correctly on the operating table. The patient and operator must
not come in contact with any metal conductive surfaces.
•
The Patient Plate must securely contact a vascular area close to the operating site. For a
gynecology procedure the preferred sites are the patient's thigh (disposable adhesive pads) or
under the patient's buttocks (reusable metal plate). The contact area must be clean, free of
body lotions, shaved, and massaged for good circulation. The contact area of the Patient Plate
must be maximized and frequently checked for uniform contact during the procedure, especially
if the patient has moved or if liquids have contacted the Patient Plate. A CONDUCTIVE GEL IS
RECOMMENDED. The Patient Plate MUST NEVER be placed so as to allow the patient's heart
to be in the pathway from the active electrode.
•
Power delivery to the operative site may be decreased appreciably if alternate pathways exist;
for example, through the metal operating table, crossed Handpiece/Patient Plate Cables, etc.
Figure 2
4
1
2
12
3
5
6
7
10
11
9
8
#11 – Suction Control
Knob Intensity
#12 – Power
Smoke Evacuator Controls
2
1
12
4
3
Summary of Contents for LEEP System 1000
Page 4: ...This page left blank intentionally 2 ...
Page 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Page 36: ...Leerseite 34 ...
Page 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Page 100: ...Page intentionnellement vierge 98 ...
Page 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Page 164: ...Página intencionalmente em branco 162 ...
Page 195: ......