Maintenance
7-13
6. Grasp the edge of the lamp-unit and carefully pull it out of the scanner.
7. Reinsert the new lamp-unit. Rest the power-connector side on the slider and carefully
push the lamp-unit into the scanner.
8. Insert the lamp-unit all the way so the internal power connections engage with the
connections on the lamp-unit.
9.
Replace the lamp cover - fit the hinges in the holes and snap the cover shut. The
cover must always be closed before turning scanner power ON.
10. Test the replacement by turning power on to see if the lamp lights.
To replace the lamp-unit:
Locate the lamp cover on the left side of the
scanner.
Open the lamp cover as described above. Pull out the
lamp-unit and replace with a new one.
Replacing the Scanner Lamp-unit:
On Scanners:
1.
Turn the scanner power off and disconnect the power plug.
2. The lamp cover is on the top right edge of the scanner.
3. Grasp the cover on each side with both hands and pull back to detach.
4. Lift the cover upwards to free it from the scanner.
5. You must replace the whole lamp-unit. A lamp-unit consists of a lamp mounted in a
reflector brace and fixed power connectors.
6. Grasp the edge of the lamp-unit and carefully pull it out of the scanner.
7. Reinsert the new lamp-unit. Rest the power-connector side on the slider and carefully
push the lamp-unit into the scanner.
8. Insert the lamp-unit all the way so the internal power connections engage with the
connections on the lamp-unit.
9.
Replace the lamp cover on the scanner - fit the bottom notches in the holes and
snap the cover shut.
The cover must always be closed before turning scanner power
ON.
10. Test the replacement by turning power and scanner on to see if the lamp lights.