Maintenance
7-11
1.
Turn the scanner power off and disconnect the power plug.
2. Place your fingers in the insertion slot and flip open the scanning area lid.
3. The drive system automatically lifts up.
4. Grasp the white pressure-platen on its braces – back edge. Tip up the back edge of the
pressure-platen and lift it out.
5. Replace with a new pressure-platen. Insert the new pressure-platen. First place the
pressure-platens front braces so they rest on the front drive
axle.
6. Tip the platen backwards until the pressure-platen’s back braces rest on the backmost
drive axle.
7. Close the scanning area lid.
NOTE:
Be sure to reset the part replacement warning flag though WIDEsystem’s “Status” tab.
Grasp by the braces on the pressure platen’s back edge.
Flip up the back edge and lift out the pressure-
platen. When you reinsert a new pressure-platen:
position so the braces rest on the parallel drive axels.
Drive Axles
The white background and pressure platen are the same. To remove the pressure-platen:
Front brace.
Back brace.
Replacing the White-background-platen:
On Scanners:
1.
Turn the scanner power off and disconnect the power plug.
2. Push down on the two lever buttons found near the insertion slot on each side of the
scanning area lid.
3. The lid’s locking mechanism should now be disengaged.
4. As you hold both lever buttons down, place your free fingers in the insertion slot and flip
the scanning area lid open.
5. On the lid’s left edge, you will find a red strip. Remove the red strip to expose the edge
of the white-background-platen.
6. Grip the edge of the white-background-platen with your fingers and pull to slide it out of
the scanner.
7. Slide in a new white-background-platen.