![Canon CXDI-55C Service Manual Download Page 98](http://html.mh-extra.com/html/canon/cxdi-55c/cxdi-55c_service-manual_61476098.webp)
2. Installation
82
<
Reference 2>
Example when the graph does not form a straight line
In the data and graph below, the data was obtained by using the Kodak MLP 190 to print
out the SMPTE test image, and the image data was measured.
(1): Shows normal data.
(2): Shows the case when the CXDI printer parameters are not set linearly.
(3): Shows the case when the printer was not calibrated properly.
SMPTE Density
Step
1
2
3
4
5
6
7
8
9
10
11
Percentage
0
10
20
30
40
50
60
70
80
90
100
0.17 0.46 0.75 1.03 1.36 1.66 1.98 2.29
2.59
2.90
3.20
(2)
Inadequate parameter settings
0.17 0.29 0.45 0.61 0.80 1.02 1.26 1.59
2.02
2.53
3.19
(3)
Inadequate calibration
0.17 0.33 0.55 0.81 1.08 1.36 1.66 1.97
2.28
2.59
2.94
(1) Normal
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1 2
3
4
5
6
7
8
9
10
11
Step
Density
Normal
Inadequate
settings
Inadequate calibration
* For improper parameters, the printer parameter “-S CS000” was not entered.
* For the inadequate calibration, the printer calibration data was set too low.
Summary of Contents for CXDI-55C
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...