2. Installation
80
4-5) As the step (3)-6), create a graph based on the data measured in step 4), and make sure
that the data from the minimum density to the maximum density nearly form a straight
line.
Characteristics of the printer may prevent the data from forming a straight line near
the minimum and maximum densities. Rotate the image on the QA screen, and reprint
or retake measurements.
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
Step
Density
Ideal Density Data for Linear Line
Step 1 2 3 4 5 6 7 8 9 10 11
3.2
D
0 0.10 0.21 0.31
0.41
0.52
0.62
0.72
0.83 0.93 1.03
3.1
D
0 0.10 0.20 0.30
0.40
0.50
0.60
0.70
0.80 0.90 1.00
12
Bit
0 132 264 396
528
660
792
924
1056 1188 1320
8
Bit 0 8 16 24
32
41
49
57
65 74 82
Step 12 13 14 15 16 17 18 19 20 21 22
3.2
D 1.14 1.24 1.34 1.45
1.55
1.65
1.75
1.86
1.96 2.06 2.17
3.1
D 1.10 1.20 1.30 1.40
1.50
1.60
1.70
1.80
1.90 2.00 2.10
12
Bit 1453 1585 1717 1849
1981
2113
2245
2377
2509 2641 2774
8
Bit 90 98 106 114
122
131
139
147
155 164 172
Step 23 24 25 26 27 28 29 30 31 32
3.2
D 2.27 2.37 2.48 2.58
2.68
2.79
2.89
2.99
3.10 3.20
3.1
D 2.20 2.30 2.40 2.50
2.60
2.70
2.80
2.90
3.00 3.10
12
Bit 2906 3038 3170 3302
3434
3566
3698
3830
3962 4095
8
Bit 180 188 196 205
213
221
229
238
246 255
Max setting
density
Film
base
density
Summary of Contents for CXDI-55C
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...