4. Repair Guide
4
2.3 Replacing the sensor cable
1) Disconnect the connector of the sensor cable which is connected to the J2 connector on the
LED board.
Do not pull the wires to disconnect the connector. Using needle-nosed pliers, push the
three positions uniformly indicated by the red arrows in the picture below. Be careful
when removing the connector so as to avoid damaging the connector.
After disconnecting the connector, check that the IC soldered to the LED board is not
damaged.
2) Disconnect the connector of the sensor cable which is connected to the J1 connector on the
LED board.
Do not pull the wires to disconnect the connector. Using needle-nosed pliers, push the
three positions uniformly indicated by the red arrows in the picture below. Be careful
when removing the connector so as to avoid damaging the connector.
After disconnecting the connector, check that the IC soldered to the LED board is not
damaged.