4. Repair Guide
3
2.2 Attaching the access cover
1)
Check that the rubber seal, wiring and cable protection sheet have been placed in their
prescribed positions.
2)
Install the access cover in such a way that it does not pinch or sandwich the rubber seal,
wiring or cable protection sheet.
3)
Attach the six mounting screws (M3x6mm) for the access cover.
Notes:
1.
No foreign matter must be allowed to enter.
2.
Be careful not to dislodge the rubber seal.
3.
Take care not to pinch or sandwich the rubber seal, wiring or cable protection sheet.
4)
Check to see that the sensor cable holder is not pinched, sandwiched or unsecured and that
there are no gaps.
Rubber seal
Cable
protection
sheet
Wiring
The sensor cable holder must not be pinched, sandwiched or
unsecured and there must be no gaps.