2. Installation
2
3 Caution on Installation
(1)
Do not install the imaging unit near electronic devices as noises and artifacts tend to appear on
images in the electromagnetic field.
e.g. CRT monitor, X-ray generator, and any other medical electronic devices.
(2)
Follow the following steps to detach the sensor cable while the system is running.
1)
To detach the sensor cable
-
Check the LED of the imaging unit and the display of the control computer to make
sure the status of communication between the imaging unit and control computer is
idle. (*1)
-
Turn off the main power switch of the power box or the remote switch. (*2)
-
Make sure the LED on the imaging unit, switch on the power box and remote switch
are turned off. Power supply to the Imaging unit must be disconnected.
-
Detach the connector of the sensor cable.
*1: Do not detach the sensor cable during the data transmission between the imaging
unit and control computer, it may cause data loss, system error or equipment
breakdown.
*2: Do not detach the sensor cable when the power is being supplied from the power box,
it may cause equipment breakdown. If you disconnect the imaging unit by improper
steps, the “Error” LED on the power box and remote switch will be turned on to alert
you the sensor cable is detached improperly. To restore from the error, turn off the main
switch of the power box or the remote switch. Then connect the sensor cable again by
following the steps described in the next column.
2)
To attach the sensor cable
-
Make sure the LED on the imaging unit, switch on the power box and remote switch
are turned off. Power supply to the Imaging unit must be disconnected.
-
Connect the connector of the sensor cable.
-
Turn on both the main switch of the power box and remote switch.
(3)
Before proceeding with installation, ensure that the static accumulated in the bodies of the
installation personnel is discharged. Similarly, before touching the PCBs (when removing
them) or cable connectors, ensure that all static is discharged.
(4)
With the launch of CXDI-60C, the specification of the CXDI-60G sensor cable is changed to
conform to CXDI-60C. For details, refer to “7.1.4 Interchanging a Imaging Unit”.
Summary of Contents for CXDI-55C
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...