2. Installation
14
(2) Cable
connections
1)
Loosen the skin top spiral and lock nut of the remote cable and X-ray I/F cable. Do not
remove the insulation lock from the X-ray I/F cable as it is put on the cable to prevent the
cable from falling off.
*The X-ray I/F cable is 20m in length. If the cable is too long, you may need to make it shorter.
Refer to “7.1.7. Adjusting the length of X-ray I/F cable” for details.
2)
Get the connector and lock nut of the remote cable through the hole in a power box.
Remote Cable
Lock nut
Skin top spiral
X-ray I/F Cable
Insulation lock
Adjust the direction of the connector
to the notch of a hole.
Get the lock nut through
the hole.