2. Installation
17
9)
Align the red marks on the connector of the sensor cable on both sides and joint the
connectors together until it locks. It snaps when it locks.
*1 Turn off the power of the power box before connecting or disconnecting the sensor
cable.
*2 Be careful not to drop the connector of the cable, it may cause injury or get the things
damaged as it is heavy.
[Improper connection]
[Proper connection]
Gap
No gap
Connector on the
imaging unit side
Connector on the
power box side
Summary of Contents for CXDI-55C
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...