2. Installation
22
7.1.5
Sensor Cable (Optional)
(1)
Sensor cable S70-60 (imaging unit side)
You can replace the sensor cable on the imaging unit side with the shorter type of 0.7meters
long sensor cable.
(2)
Sensor cable SP780-60 (Straight type)
The sensor cable can be replaced with the straight type sensor cable with no detachable
connectors.
(3)
Sensor cable P630-PM (Panel mount type)
This sensor cable can be used in place of a standard sensor cable P630 (power box side). The
connector of the sensor cable P630-PM(panel mount type) can be tightened to a stand or a table.
Note that an insertion hole of 24.1mm in diameter is necessary for a stand or a table to fasten the
connector.
Ethernet
Ethernet
Power box
Operation unit
Control
Computer
X-ray tube
X-ray
generator
CXDI-60G/60C
Imaging unit
Summary of Contents for CXDI-55C
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...