6. Troubleshooting
14
6.2.2 When problems occur in CXDI software
When problems occur in files other than ini files for d: \Ccr software other than the OS, recovery
can be accomplished by adding or replacing the relevant files if you can identify the defective
files. Recover Ccr with the following procedure if you cannot identify the defective files:
1) Copy d: \ccr to a different directory or laptop computer.
2) Delete d: \ccr, and then newly install the CXDI software. (Note)
(Note) The same version of the software must be installed.
3) Copy the following files in Ccr that were copied in Step 1) to d: \ccr
\dtstore (captured images)
\Logs
\White#.dcm
\defpix#.dat
\xxxxxx.dp
4) Newly install the optional software (DMW/Gen.communication) if you are using it.
5) Copy to overwrite all the ini files in d: \Ccr stored on the floppy disk in order to recover the
user’s environment.
6) Perform the settings again that are described in the “Control PC serial number” and “Setting
the imaging unit identification and the number of connecting units” in this manual “Setting
Procedures”.