Canon CXDI-55C Service Manual Download Page 1

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

C

C

X

X

D

D

I

I

-

-

6

6

0

0

G

G

/

/

6

6

0

0

C

C

 

 

C

C

a

a

n

n

o

o

n

n

 

 

I

I

n

n

c

c

.

.

 

 

J

J

a

a

p

p

a

a

n

n

 

 

C

C

o

o

p

p

y

y

r

r

i

i

g

g

h

h

t

t

 

 

(

(

C

C

)

)

 

 

C

C

a

a

n

n

o

o

n

n

 

 

I

I

n

n

c

c

.

.

 

 

M

M

e

e

d

d

i

i

c

c

a

a

l

l

 

 

T

T

e

e

c

c

h

h

n

n

i

i

c

c

a

a

l

l

 

 

S

S

e

e

r

r

v

v

i

i

c

c

e

e

 

 

D

D

e

e

p

p

t

t

.

.

 

 

A

A

l

l

l

l

 

 

r

r

i

i

g

g

h

h

t

t

s

s

 

 

R

R

e

e

s

s

e

e

r

r

v

v

e

e

d

d

.

.

 

 

S

S

e

e

r

r

v

v

i

i

c

c

e

e

 

 

M

M

a

a

n

n

u

u

a

a

l

l

 

 

P

P

r

r

i

i

n

n

t

t

e

e

d

d

 

 

b

b

y

y

 

 

C

C

a

a

n

n

o

o

n

n

 

 

I

I

n

n

c

c

.

.

 

 

J

J

u

u

l

l

.

.

 

 

2

2

0

0

0

0

9

9

 

 

R

R

e

e

v

v

.

.

0

0

1

1

 

 

Confidential

Summary of Contents for CXDI-55C

Page 1: ...Ca an no on n I In nc c M Me ed di ic ca al l T Te ec ch hn ni ic ca al l S Se er rv vi ic ce e D De ep pt t A Al ll l r ri ig gh ht ts s R Re es se er rv ve ed d S Se er rv vi ic ce e M Ma an nu ua al l P Pr ri in nt te ed d b by y C Ca an no on n I In nc c J Ju ul l 2 20 00 09 9 R Re ev v 0 01 1 Confidential ...

Page 2: ...notes If the product undergoes a large modification a revised edition of the service manual will be sent to you In other cases a service manual report will be sent to you to update the manual Note 1 This service manual is published by Canon Inc in accordance with Article 6 Furnishing the Referring Materials of the Service Assignment Contract it has concluded with your company Note 2 This service m...

Page 3: ... cover during maintenance repair etc perform the work after switching the power off Never touch the device with wet hands as there is a risk of electric shock 3 Fuse When replacing the fuse first resolve the reason for its failure and then replace the fuse with the specified type Never use a fuse other than the specified type 4 Connecting the grounding wire The provided ground wire must be connect...

Page 4: ...ung des Instruments mit nassen Händen Anderenfalls können Sie einen elektrischen Schlag erleiden der zum Tod oder schwerer Verletzung führen kann 3 Sicherung Wenn die Sicherung ausgewechselt werden muß schalten Sie unbedingt die Stromversorgung des Instruments aus und beheben Sie die Ursache für das Durchbrennen der Sicherung Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ Anderenfal...

Page 5: ...cited from IEC60601 1 1 2000 the measurements are only guidelines However the IEC60601 1 1 2000 example must be treated as the standard Therefore the Control PC and operation unit must be installed in a location further than the measurements below outside of the patient environment Areas where the patient moves not only during imaging but when entering and leaving the room etc are also considered ...

Page 6: ... pa an n C Co op py yr ri ig gh ht t C C C Ca an no on n I In nc c M Me ed di ic ca al l T Te ec ch hn ni ic ca al l S Se er rv vi ic ce e D De ep pt t A Al ll l r ri ig gh ht ts s R Re es se er rv ve ed d 1 1 G Ge en ne er ra al l Printed by Canon Inc Jul 2009 Rev 01 ...

Page 7: ...Contents 1 General 1 2 CXDI System Block Diagram 2 3 System Diagram 3 4 CXDI Image Processing 4 5 Specifications 5 ...

Page 8: ... LANMIT Image data is read out from one direction The appearance of the CXDI 60G is almost the same as that of the CXDI 60C but their fluorescence substances that convert X ray into visible light are different CXDI 60G CXDI Software Ver 7 0 and later Serial number of CXDI 60G imaging unit 100001 and later Serial number of CXDI 60G imaging unit non logo model 300001 and later CXDI 60C CXDI Software...

Page 9: ...peration Unit RS232C VGA Network Ethernet 100 10base T AC100V 240V 10 50 60Hz 2 2A Mouse Power Box OUT PUT 1 Control Signal Power OUT PUT 2 Control Signal AC Power IN AC100 120V 230V CA1 CA6 CA7 CA8 Imaging Unit CXDI 60G 60C Control Signal Power Control computer VGA LAN1 AC Power IN MOUSE KEY BOARD Key Board Printer Image file device Image diagnosis W S X Ray I F CA2 LAN2 CA3 CA4 CA5 Remote Switch...

Page 10: ...dy been connected Extend the ethernet port by general switching HUB or ethernet card The maximum number of connections is limited to four by the control software specification Total System Block Diagram Example AC or DC IN Protective Grounding 1st 2nd AC250V Insulation with reinforcement Power Supply AC250V Basic Insulation Imaging Unit Grid Unit Remote Switch X ray generator 601 Compatible FC 24V...

Page 11: ...processing The CXDI performs image processing up to this point 5 DIAGNOSIS IMAGE QA image after further image processing necessary for diagnosis Image processed by the user for diagnostic purposes 6 PROCESSED IMAGE Diagnosis image after post processing Image modified by the user or the default processed image Born image Raw image Original image QA image Diagnosis image Processed image Correction p...

Page 12: ...r Through the Power Box Imaging time 1 sec Imaging cycle 15 sec Dimension 324 W x 327 D x 20 8 H mm 344 4 W x 380 D x 22 5 H mm Imaging Unit coloring Art gray Cool white Imaging Unit mass w o grid 2 8Kg Except the cable 3 3Kg Including the cable 2 5Kg Except the cable 2 8Kg including the cable Space between surface where patient gets in contact CFRP and sensor surface glass 5 75mm Within 5 1 0 5mm...

Page 13: ...the internal temperature is below 44 degree Celsius When the internal temperature of Imaging Unit is above 49 degree Celsius its state is changed to sleep mode And the Imaging prohibition will be continued when the internal temperature is below 48 degree Celsius User interface Indication Indication Busy Sensor Power Color Two colors Color Orange 1 Green Blue Imaging unit is off Off Imaging unit is...

Page 14: ...ngth 10bit Data rate 15 625 kHz Power supply AC 100 240V 50 60Hz 1 2 0 7A Reference 60 Sensor Power supply Rated Voltage AC 100 240V AC 85 264V Mass 3 7 Kg Except the external cables Dimension 358 W x 200 D x 65 H mm Except bottom rubber parts With bottom rubber parts 75mm 3 Environment rated parameters Item Content Remarks Ambient temperature 5 to 35 C Relative humidity 30 to 75 RH Operating envi...

Page 15: ...D DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 2 2 I In ns st ta al ll la at ti io on n M Ma an nu ua al l Printed by Canon Jul 2009 Rev 01 ...

Page 16: ... Using the Imaging Unit in the Multiple Rooms 19 7 1 4 Interchanging a Imaging Unit 20 7 1 5 Sensor Cable Optional 22 7 1 6 Connection Diagram for Control Computer Rear Panel 23 7 1 7 Adjusting the Length of the X ray I F Cable 24 7 2 Starting up and Shutting down the System 25 7 2 1 Sequence for Starting up the System 25 7 2 2 Sequence for Turning the Power off 25 7 3 X ray Controller Interface 2...

Page 17: ... an angle or subjected to vibration or shock this includes during transportation 4 The equipment must not be kept where chemical products are stored or where gasses are generated 5 The site s power supply must be of the correct voltage and frequency for the equipment 6 The site must be connected to a fully earthed cable with sufficient ground resistance to meet standard values 3 After installation...

Page 18: ...wn 2 Do not detach the sensor cable when the power is being supplied from the power box it may cause equipment breakdown If you disconnect the imaging unit by improper steps the Error LED on the power box and remote switch will be turned on to alert you the sensor cable is detached improperly To restore from the error turn off the main switch of the power box or the remote switch Then connect the ...

Page 19: ... No Item Name Qty Remarks 1 Sensor Cable P630 PM Panel mount type 1 6 3m 2 Sensor Cable P70 60 imaging unit side 1 0 7m 3 Sensor Cable SP780 60 Straight type 1 7 8 m LAN cable for connecting Control Computer Power Box and Network switch Switching HUB for connecting the multiple Imaging Units shall be procured at each sales company LAN cable Over category 5 Recommended length of the cable is 30m or...

Page 20: ...s Sensor cable S150 60 imaging unit side Remarks 6 3m No 3 No 4 Name Power Box Name X ray I F cable Qty 1 Qty 1 Remarks I F and Power supply Remarks Connection with X ray generator No 5 No 6 Name Remote switch Name Power supply cable Qty 1 Qty 1 Remarks Switch to turn on and off Power Box Remarks For Power Box 100 120 230V each type ...

Page 21: ...he difference between them is the form of the connector There is no difference in the function of the cables Sensor cable S70 60 imaging unit side This sensor cable can be used in place of a standard sensor cable S150 60 imaging unit side Only the difference between them is the length of the cable The sensor cable S70 60 is 0 7m in length There is no difference in the function of the cables Sensor...

Page 22: ...stem CXDI 60G 60C 1 CXDI 60G 60C Imaging Unit Package Sensor cable P630 6 3m Corner pad Imaging unit Corner pad Center pad Paper tube Place the sensor cable S150 60 on the packet corner of the paper tube Positioning sheet Cardboard sheet Cardboard box for imaging unit ...

Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...

Page 24: ... 1 Outer packaging 2 Inner packaging Material label Corner Packing Outer package Handle with care label Origin label Shipping environment label Product division label Grid unit Plastic bag Inner package Protection board Protection sheet ...

Page 25: ...icrosoft Windows XP Professional recommend 3 LAN Card 1 For Laptop computer as required 4 Mouse 1 PS 2 type 5 Keyboard 1 PS 2 type 6 HUB 1 Connection between Control Computer and Laptop Computer 7 10 100BASE TX cable 2 Straight type Control PC to Laptop computer 8 CXDI Software version compatibility table 9 Mirror oil based marker and paper etc 1 For adjusting the alignment with the X ray tube ...

Page 26: ...ensor serial numbers 6 Identifying the Imaging Units in section 7 6 8 Enter control computer serial number 7 Entering Control Computer Serial Number in section 7 6 9 Adjusting the timing with X ray generator No required usually 10 Calibration No error must be displayed Operation Manual 11 Setting the Fixed ROI Areas If necessary set the ROI area 12 Set exposure parameter table Set it in consultati...

Page 27: ...sulted prior narrowing down the adjustments to the final values Operation Manual 25 Inserting the backup floppy disk It must be confirmed at re start that backup files have been made No necessary for the system installed in vehicles 15 Backing up Setting Data to FD in section 7 6 26 Back up valuable data Copy the CCR folder to the removal drive 14 Backing Up when Installing in section 7 6 ...

Page 28: ...2 Installation Manual 12 7 Installation 7 1 Connection to Each Unit 7 1 1 Connection Diagram CXDI 60G 60C POWER BOX PWB XRAY POWER SUPPLY J4 J2 J3 J1 Control computer Remote Switch ...

Page 29: ...t the bottom of the power box to remove the top cover Different types of screw are used for backside and lateral side Make sure to use the proper type of screw respectively on installation 2 Before connecting the X ray I F cable and remote cable remove the LAN cable connector guard to avoid injury Also remove screws on the far side LAN cable connector guard ...

Page 30: ...from falling off The X ray I F cable is 20m in length If the cable is too long you may need to make it shorter Refer to 7 1 7 Adjusting the length of X ray I F cable for details 2 Get the connector and lock nut of the remote cable through the hole in a power box Remote Cable Lock nut Skin top spiral X ray I F Cable Insulation lock Adjust the direction of the connector to the notch of a hole Get th...

Page 31: ...ngth for wiring After adjusting the length properly fasten the skin top spiral tightly Note To avoid the risk of damage when the cables are removed with very large force check skin top spiral of the bush refer the following figure Remote cable X RAY I F cable 4 Fasten the skin top spiral and lock nut tightly using 15mm screw wrench Gap 1 0 to 1 5mm No gap Skin top spiral Lock nut Connector ...

Page 32: ...ower box and LAN cable connector guard Different types of screw are used for backside and lateral side Make sure to use proper type of screw respectively on installation 8 Connect the sensor cable LAN cable AC cable to the backside of the power box 1 Push the connector of the sensor cable into the connector terminal and fasten it tightly until the lock ring stops 2 Only the AC cable attached to th...

Page 33: ...It snaps when it locks 1 Turn off the power of the power box before connecting or disconnecting the sensor cable 2 Be careful not to drop the connector of the cable it may cause injury or get the things damaged as it is heavy Improper connection Proper connection Gap No gap Connector on the imaging unit side Connector on the power box side ...

Page 34: ...he lock and pull out the connector of the imaging unit side straightforward Do not grab the cable itself 1 Turn off the power of the power box before connecting or disconnecting the sensor cable 2 Be careful not to drop the connector of the cable it may cause injury or get the things damaged as it is heavy Release the lock ...

Page 35: ...e rooms with the CXDI SYSTEM II installed by the service engineer The steps to connect and disconnect the detachable connector of the sensor cable are described in the operation manual Imaging Unit X Ray Room 2 Generator Control Computer Power Box Sensor Cable X Ray Room 1 Generator Control Computer Power Box Sensor Cable ...

Page 36: ... old type sensor cable plug and socket of the connector does not match Therefore you need to replace it with the new type sensor cable With the launch of CXDI 60C the new type sensor cable which is conform to CXDI 60C has been adopted to CXDI 60G Ethernet Ethernet Power box Operation unit Control Computer X ray tube X ray generator CXDI 55G 55C imaging unit CXDI 60G 60C imaging unit ...

Page 37: ... connector of the old type sensor cable on the CXDI 60G imaging unit side For the CXDI 60G imaging unit having the old type sensor cable plug and socket of the connector does not match Refer to the Service Manual report CXDI 60G 09 002 for details CXDI 55G 55C Imaging Unit Power Box CXDI 60G 60C Imaging Unit Power Box Power Box CXDI 60G Imaging Unit Old type sensor cable ...

Page 38: ...e sensor cable with no detachable connectors 3 Sensor cable P630 PM Panel mount type This sensor cable can be used in place of a standard sensor cable P630 power box side The connector of the sensor cable P630 PM panel mount type can be tightened to a stand or a table Note that an insertion hole of 24 1mm in diameter is necessary for a stand or a table to fasten the connector Ethernet Ethernet Pow...

Page 39: ...S 232C cable to the COM port COM 1 or COM2 specified as the port in the driver for Operation Unit LAN 2 For connecting to a power box LAN 1 For connecting to a network Keyboard Mouse Power Supply VGA Port For Operation Unit COM1 RS 232C Port For Operation Unit COM2 RS 232C Cable For Operation Unit ...

Page 40: ...Electrical specification Photo coupler Current loop with the resistance 100 Ω or less Retardation Retardation time 1msec or less Parts Key No Part number Description Remark 1 Y67 2868 000 Splice terminal V0 5 4 Clear JST Tool number YNT 2622 JST Y67 2869 000 Wire Marker Mark A Y67 2870 000 Wire Marker Mark B Y67 2871 000 Wire Marker Mark 1 2 Y67 2872 000 Wire Marker Mark 2 VS 2 HAGITEC 3 KE2 2101 ...

Page 41: ...g on the control computer Since the power box is equipped with a remote switch that turns on off the secondary output you can install the switch on your side to turn it on off 1 Turn on the main power of the power box 2 Turn on the remote switch of the power box 3 Turn on the control computer Note Ccrstart bat should be registered in Windows Startup 7 2 2 Sequence for Turning the Power off Shutdow...

Page 42: ...ission End X ray emission X ray emission signal OFF Detect X ray emission signal ON Imaging possible Enable X ray emission ON X ray Generator CXDI Image Data read RX_REQ Generator Setup timer X ray emission signal 1st SW level ON X ray emission signal 2nd SW level ON RX_COM Enable X ray emission OFF ...

Page 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...

Page 44: ...OM is time out as a trigger reading action is delayed if RX_COM does not negate and time out is used as the trigger resulting in delay of image display timing We recommend using a configuration in which RX_REQ is negated Time required from X ray exposure completion to negating RX_REQ TRX_REQ_NEG min 0 max 100ms Image display timing is delayed if this is not fulfilled RX_COM X ray release signal Ch...

Page 45: ...R F NC2D P DC5V The connection with CXDI 60G 60C and X ray generator equipment Connection conditions 1 The X ray exposure signal line including the switching function must be insulated and its total impedance must be 100 ohms or less 2 The maximum contact voltage of the X ray exposure authorization signal line is AC 250V and DC 30V and its current ranges from 10mA to 2A Only the insulated secondar...

Page 46: ...le arrangement 2 Form C Contact material Au clad AgNi type Relating capacity AC250 5A DC30V 5A Max switching power 1250VA 150W Max switching voltage 250V AC Max switching current 5A Min switching capacity 100µA 1V DC 3 Characteristics Item Content Operate time Max 20ms Reset time Max 10ms Mechanical 18 000 times hour Maximum open close frequency Rated load 1 800 times hour Between coil contacts 2 ...

Page 47: ...tage VF IF 10 mA 1 0 1 15 1 3 V Reverse Current IR VR 5 V 10 µA LED Capacitance Ct V 0 f 1 MHz 30 pF Collector Emitter Breakdown Voltage V BR CEO IC 0 5 mA 80 V Emitter Collector Breakdown Voltage V BR ECO IE 0 1 mA 7 V VCE 48 V Ambient Light Below 100lx 0 01 2 0 1 10 DETECTOR Collector Dark Current ICEO VCE 48 V Ta 85 Ambient Light Below 100lx 2 4 50 50 µA Rise Time tr 2 Fall Time tf 3 Turn On Ti...

Page 48: ...on Set up the following three items 1 TCP IP setting for the control computer network card 2 Network setting for Screwcap ini 3 Network setting stored in the imaging unit 2 Preparation Keyboard Mouse Connect the keyboard and mouse to the back of the control computer Check if the system is connected and then turn the system on 3 Setup method 3 1 TCP IP setting for the control computer network card ...

Page 49: ...d that communicates with the imaging unit in the previous item 3 1 TCP IP setting for the control computer network card Screwcap ini is located in the following directory D ccr screwcap ini 1 Imaging Unit IP address This address must be the same as the IP address stored in the next item 3 3 Network setting stored in the sensor HostInfo ResponseTimeout 30000 DataTimeout 60000 IntervalTimeout 30000 ...

Page 50: ...sary unless you have changed the setting Item to be set Factory default value Imaging unit IP address 192 168 100 11 Subnet mask 255 255 255 0 Host IP address 192 168 100 10 Port number for command 12121 Port number for data 12122 If you change the setting refer to Tool Software Operation Manual for Ethernet Imaging Part IP Address Setting ...

Page 51: ...utomated Exposure Control Field in Preview Screen added from CXDI System Software Ver 6 2 NUMBER that can be specified is max 3 Ws 234mm Hs 284mm 0 0 W H Sensor X Y 2 3 Convert the size and position of the ROIs in 2 to pixel values The pixel size of the sensor is 160µm For multiple values use X Y W H X Y W and H for calculations X 160 µm Let this value equal A Y 160 µm Let this value equal B W 160...

Page 52: ...ton Press the Fixed ROI Area key 2 6 Press the Area setting key 2 7 The fixed ROI 1 setting screen appears Input values A to D from step 3 into the edit box and press ENABLE To set multiple fixed ROIs input A to D and A to D into fixed ROI 2 settings and fixed ROI 3 settings respectively Area setting button A A B B C C D D ...

Page 53: ... dimmed press the Area setting key and press DISABLE in fixed ROI settings 2 9 After Confirming all settings and press EXIT 2 10 The display returns to the system settings screen Press OK 2 11 Change settings appears Press OK Be careful because if CANCEL is pressed all changes made to the settings are deleted 2 12 Return to the normal imaging screen and turn off the power to the CXDI ...

Page 54: ...The date and time is set to Japan standard time at factory shipment Reset the date and time to your local value as necessary 2 Procedure 2 1 When CXDI application start open the ADMINISTRATOR SETUP MENU SYSTEM SETUP MENU ADMINISTRATOR SETUP 2 2 When the ADMINISTRATOR SETUP MENU appears and presses the DATE button 2 3 The dialog Date Time Properties appears and set the value properly each of the fi...

Page 55: ... information from the user mode SYSTEM SETUP MENU VERSION INFORMATION c Confirm the firmware version System Module Version Information A Hardware version Product type Product type is identified from Sensor serial No that is set to the Imaging unit 0b indicates 60G and 0f indicates 60C B Firmware initialization code version This is the version of the initialization code written on the PWB 60Di Init...

Page 56: ...to Telnet Connection in the Tool Software Operation Manual for Ethernet 5 Check the versions of the firmware and PLD code on the screen displayed after the login 6 After you finish checking close HyperTerminal login canon Password CXDI 60 firmware Ver 3 00 08 build as May 19 2008 10 39 06 normal boot HUB60 Ver 6004 prod code 0bc0 pwb 60di Firmware version Hardware version PLD code version ...

Page 57: ...CXDI is connected to the system 3 Procedure 3 1 Installing the firmware Write the firmware by referring to Firm Write Tool Software Firmwrite exe in the Tool Software Operation Manual for Ethernet Where to write PWB 60Di 3 2 Installing PLD code Write PLD code by referring to PLD Write Tool Software pldwrite exe in the Tool Software Operation Manual for Ethernet Where to write PWB 60Di ...

Page 58: ...er the CXDI software is launched and an error message appears In that case you need to register the sensor serial number 2 Notes 2 1 Check the sensor serial number whenever the imaging unit is replaced 2 2 This checking procedure must be performed with the Control computer Imaging Unit and all the other equipments connected and started up 3 Sensor serial number registration Register the sensor ser...

Page 59: ...Startup Group 3 Register the CXDI software to the Startup Group procedure 3 1 Connect keyboard and mouse to the control computer 3 2 Turn the all CXDI system power on after the all installation finished And after that Windows XP starts 3 3 Open the Taskbar and Start Menu from the Start Menu Start Settings Taskbar and Start Menu 3 4 Taskbar and Start Menu Properties appears Click Start Menu tab and...

Page 60: ...Buffer Size to 5000 Click OK 4 6 The Apply Properties to Shortcut appears and check the item of the Modify shortcut which started this window Click OK 5 Delete CXDI software from the Startup Menu Group 5 1 Connect the keyboard and the mouse to the control computer 5 2 Turns the all CXDI system power on Windows XP start 5 3 After the CXDI software start press Alt Tab key to show the CCR Console Men...

Page 61: ...g file appears If you are going to remove it click Yes button 5 10 After confirm that the ccrstart bat item is removed from Startup Group close all the application on the desktop and re login to Windows XP 5 11 Make sure that the CXDI software will not start automatically after login to Windows XP And then shutdown Windows XP turn the CXDI system power off When the CXDI software is deleted from th...

Page 62: ... 4 Setting Mode 0 Normal 1 Expert 0 0x0 appears Select 0 Normal 4 5 CCR SETUP MENU appears Select 7 Scan Sensor Setup 4 6 The Capture Device Configuration Table appears Input 1 in Max Capture Devices shown below 4 7 Enter the serial number to A D Board Serial Number for SensorID 1 Capture Device Configuration Table Max Capture Devices 1 0x1 1 Capture Device Configuration No 0 SensorID 1 OPU A D Bo...

Page 63: ... 0 0 0 is prompted Select 0 Normal 2 5 CCR SETUP MENU appears Select the command 1 System Setup 2 6 CCR Serial Number 0 0 0 appears Enter the six digit number indicated on the naming label of the control computer unit Press Enter key until CCR SETUP MENU appears Fig 1 Fig 1 2 7 Press Esc key after CCR SETUP MENU appears to return to Welcome to CCR screen 2 8 Select 8 Exit to exit CXDI software 2 9...

Page 64: ...re 2 2 Open the TABLE SETUP Change window from the Normal Exposure window System SETUP MENU SYS SETUP TABLE SETUP 2 3 Select the tabs to be changed and change the X ray tube voltage X ray tube current and msec value data to match the exposure conditions of the X ray generator See the operation manual for the details of settings 2 4 After finishing the changes return to the Normal Exposure window a...

Page 65: ...ed for the annotation are performed 2 Procedure 2 1 Once the normal radiographic screen has started open the annotation setting screen SYSTEM SETUP MENU SYS SETUP ANNOTATION 2 2 The annotation setting screen now appears Proceed with the settings that will make it possible to put the data desired by the user See the operation manual for the details of settings ...

Page 66: ...fully since any errors made in these settings will make it impossible for connection to be made to the network or the images to be transmitted properly etc 3 Windows XP settings 3 1 Connect the keyboard and mouse to the control computer 3 2 After turning on the Operation unit s power and then the Control computer s power start Windows XP 3 3 The Windows XP desktop screen appears Right click the My...

Page 67: ...nstance the following messages will be repeated y If the CXDI has been connected properly Pic ping 172 17 7 123 input on the DOS screen Pinging 17217 7 123 With 32 bytes of data Reply from 172 17 7 123 bytes 32 time 10ms TTL 255 Reply from 172 17 7 123 bytes 32 time 10ms TTL 255 Reply from 172 17 7 123 bytes 32 time 10ms TTL 255 Reply from 172 17 7 123 bytes 32 time 10ms TTL 2550 If the CXDI has n...

Page 68: ...nstallation investigation details a Printer host name IP address b Port number c Transmission destination title Fig 2 C Press the SET button and input the parameters of the printer to be connected based on the pre installation inspection details Refer to another sheet for details of the parameters A space delimiter must be input between each of the parameters By pressing the Override button you ca...

Page 69: ...set B Press the Storage1 button and input the following items based on the pre installation investigation details a Storage host name IP address b Port number c Transmission destination title Fig 4 C Press the SET button and input the parameters Normally the parameters need not be set They must be input only when the need arises A space delimiter must be input between each of the parameters Fig 5 ...

Page 70: ... during the association Retry Message a indicates that connection at the TCP IP level is not possible and that the physical connections or the subnet mask and other settings must be checked again Message b indicates that communication at the TCP IP level is problem free but that DICOM level communication has failed In this case check again that AE_TITLE of CXDI has been sent properly to the transm...

Page 71: ...ling App Entity Title The CXDI automatically uses CANON_CCR internally for the operation Designating a specific value allows overwriting of the above value Note that the meaning is opposite of the DICOM printing argument c ÆSee the printing parameters This is used when the operator who manages the connected storage device requests a change in the installed identification information version s SOPN...

Page 72: ...T OD is enabled set appropriate options to suit each output destination level 0 Do not delete default when not specified 1 Delete Window Center Width 2 Delete Window Center Width and Rescale Intercept Slope Type 3 Delete Rescale Intercept Slope Type compatible with releases up to Ver 4 20 Other than a loadable LUT or γ 1 0 has been specified with the output destination specific LUT function For st...

Page 73: ...image is inserted into the queue in the DICOM printer At HIGH the image is printed first among the queued images This parameter is set according to the user s requirement In the DICOM library TYPE3 the value is transferred together with the Tag However if the value is unknown the value is either transferred as a character string with length 0 or the element itself is not transferred Æ The printer ...

Page 74: ...meter is not transferred problems can occur since unsuitable default values may be used Some printers do not print until a supply magazine of the designated size is loaded and others print even though the designated size is different from the currently loaded supply magazine M magnification Interpolation method NONE REPLICATE BIL INEAR CUBIC This parameter is used in the DICOM Basic Film Box 2010 ...

Page 75: ...ase the printer default values are used Starting from version 2 0 the image can be rotated from the CXDI side without using this parameter A max_density Maximum density Dx100 This parameter is used in the DICOM Basic Film Box 2010 0130 This parameter designates the density of the digital value for 0 4095 for reverse display of the CXDI transfer data image pixels In the CXDI 0 indicates black In th...

Page 76: ... Image Box 2020 0020 Reverse image density Æ If this parameter is not designated it is not transferred over DICOM In this case the printer default values are used r pixel_pitch Transfer pixel pitch for designating the request image size Pixel Pitch in um This parameter is used in the DICOM Basic Image Box 2020 0010 Position of the image on film The cumulative value for the horizontal size of the i...

Page 77: ...he CXDI default settings the printer information is not designated In this case the printer side sends all the information that it has DICOM official specifications When the g option is added the essential information only is collected This information includes the Printer Status and Printer Status Info To prevent installation when the printer does not satisfy the above DICOM specifications Æ Norm...

Page 78: ...r is to be changed when taking timeout k level 0 1 2 The k option has been implemented to normalize DICOM headers Note The parameters p v and V filename display the log on the console Therefore be sure to always erase these parameters before operation by the user In the past k2 was used to fix troubles but it has now been set as the default has been changed to k2 to ensure precise compliance with ...

Page 79: ...er manufacturer must be performed Kodak Imation DryView 8700 Pacs LINK IMN 9410 A 310 T NO M CUBIC m S LUT 0 2 Site Dependent entered in param member 78 entered in pixelPitch member 14x17 4096 entered in W member 5220 entered in H member Trimming OFF Cubic spline interpolation Smoothing must be adjusted at the user s site Maximum density 3 10 S LUT m n is designated as the Config Info but m and n ...

Page 80: ...in W member 4277 entered in H member Trimming OFF Cubic spline interpolation Smoothing must be adjusted at the user s site Maximum density 3 10 S LUT m n is designated as the Config Info but m and n are adjusted on site by the Kodak service engineer Basically adjustment is performed to achieve a linear output Nishimoto EL2000N A 320 T NO M CUBIC S 15 entered in param member 80 entered in pixelPitc...

Page 81: ...nsity of 3 20 cannot be designated For this reason a non linear LUT is required Konica Drypro 722 Printlink A 320 T NO M CUBIC m 2 S KC_LUT 1 O PORTRAIT P NORMAL B BLACK entered in param member 80 entered in pixel pitch member 14x17 4424 entered in W member 5324 entered in H member 14x14 4424 entered in W member 4372 entered in H member 11x14 3436 entered in W member 4424 entered in H member Trimm...

Page 82: ...er itself is 79 µm x 4090 x 5260 this complete resolution cannot be used when the image passes through the spooler When _ is used in AE Title the association is rejected Use the Disable function for N EVENT REPORT to disable this setting The FilmSize parameter can be used The MediaType BLUE CLEAR parameter is also supported Although the Film Orientation parameter is supported up to 1 7 MB in actua...

Page 83: ...ew 8700 8800 78 µm x 4096 x 5220 Note The 8700 printer cannot display in landscape orientation Also the maximum density is 3 1 If the 8800 box is not added this printer cannot be used by the CXDI However it can be used starting from CXDI version 2 0 If there is an Imager Pixel Spacing tag 0018 2264 the imager will fail 1996 x 2544 160 µm Fuji CR DPL FM DPL FM PS551 100 µm x 3520 x 4280 value after...

Page 84: ...2 Installation 68 DMAX 3 2D equivalent curve required by CXDI Gamma type 17 SAR system Density Shift Contrast Point 1 1 57 0 10 1 00 Point 2 2 29 0 15 1 00 ...

Page 85: ...T_CONNECT The log output in this case is shown below 18 135933 d2 ERR d0012 Attempt to connect to unknown host test 18 135933 d2 ERR 130012 Peer aborted Association or never connected 18 135933 d2 ERR 180012 Failed to establish association After a command request was sent to the server an error was returned in response Return value 0x10 CXDI description DICOM response error CCRTRANS_ERR_RESP The l...

Page 86: ... case is shown below Not determined Other errors Return value Values other than those above CXDI description DICOM communications error CCRTRANS_ERR_DICOM_TRANSE The log output in this case depends on the specific error A typical example is shown below 18 140933 d2 ERR c0082 SRV Send DATA SET failed in SRV_SendDataSet 18 140933 d2 ERR 190082 SRV Request failed in SRV_NCreateRequest 18 140933 d2 ER...

Page 87: ...uture point in time will be helpful Hitachi y Transmission is currently performed with the 1 study multi series 1 series 1 image setting However since a multi format is used for the screen displays for each series the switching operations are a hassle The user will find it more convenient if it is at all possible to change the setting to 1 study 1 series 1 series multi image y This problem arises ...

Page 88: ... server See below for the log output at this time To be determined Other errors Return value other than above CXDI interpretation DICOM communication error CCRTRANS_ERR_DICOM_TRANSE The log output is many and varied It depends on the error One example is shown below 18 140933 d2 ERR c0082 SRV Send DATA SET failed in SRV_SendDataSet 18 140933 d2 ERR 190082 SRV Request failed in SRV_NCreateRequest 1...

Page 89: ...es which cannot be changed by users with 15 types of characteristic curves which can be changed by users Concerning connections Two types of the Dry View 8700 are available Dry View 8700 Plus The 8700 Plus is a printer which can be connected to two diagnostic units When used in combination with the 8800 multi input manager it can be connected to up to eight diagnostic units Images are processed as...

Page 90: ... to 2 units can be connected Dry View 8700 Plus 8800 Multi Input Manager Up to 8 units can be connected Dry View 8700 Dual 8800 Multi Input Manager Up to 2 printers can be connected only with Dry View 8700 Dual Up to 7 units can be connected ...

Page 91: ...ter and monitor image output settings should be set to LINEAR For example if the KODAK MLP190 is connected the printer parameter S configuration must be set to S CS000 For other printers refer to Printers and Parameter Examples Reference in Network Connections and set so that the curve shape is 0 density linear B Be sure to calibrate the printer and high definition monitor units separately before ...

Page 92: ...0 to 100 of test image grayscale on the film or on the monitor Fig 3 Measure the SMPTE image density on the film is measured using a densitometer Measure the SMPTE image density on the high definition monitor using the gradation analysis software The data for the SMPTE test image grayscale transferred by the CXDI are the values for the maximum density 3 20 in the case of the MLP 190 in the printer...

Page 93: ...ents when the maximum density is set to 3 2 D and the printed film is measured with a densitometer Refer to the data below when the maximum density is set to 3 1 D or when gradation analysis software is used to take measurements on the monitor The LINEAR output cannot be set for some printer models In this case try to select an output setting as close as possible to LINEAR Ideal density values in ...

Page 94: ...posure adjust the trim so that the gray scale may be located in the center of the image Fig 5 For correct density measurement of the 32 step chart the gray scale must be printed in the center of an image to eliminate the shading feature Data may not from a straight line near the minimum and maximum densities due to characteristics of the printer Rotate the image on the QA screen and reprint or ret...

Page 95: ...e on the test image Fig 7 1 Measure the SMPTE image density on the film using a densitometer The SMPTE image density on the high definition monitor is measured using the gradation analysis software 2 The data for the SMPTE test image grayscale transferred by the CXDI are the values for the maximum density 3 20 in the case of the MLP 190 in the printer settings which have been changed into 32 unifo...

Page 96: ... 10 0 21 0 31 0 41 0 52 0 62 0 72 0 83 0 93 1 03 3 1 D 0 0 10 0 20 0 30 0 40 0 50 0 60 0 70 0 80 0 90 1 00 12 Bit 0 132 264 396 528 660 792 924 1056 1188 1320 8 Bit 0 8 16 24 32 41 49 57 65 74 82 Step 12 13 14 15 16 17 18 19 20 21 22 3 2 D 1 14 1 24 1 34 1 45 1 55 1 65 1 75 1 86 1 96 2 06 2 17 3 1 D 1 10 1 20 1 30 1 40 1 50 1 60 1 70 1 80 1 90 2 00 2 10 12 Bit 1453 1585 1717 1849 1981 2113 2245 23...

Page 97: ...As shown in the graph the data from the minimum density film base density to the maximum density nearly form a straight line SMPTE Density Step 1 2 3 4 5 6 7 8 9 10 11 Percentage 0 10 20 30 40 50 60 70 80 90 100 Density D 0 17 0 46 0 75 1 03 1 36 1 66 1 98 2 29 2 59 2 90 3 20 0 00 0 50 1 00 1 50 2 00 2 50 3 00 3 50 1 2 3 4 5 6 7 8 9 10 11 Step Density The printer parameter settings in this case ar...

Page 98: ... Density Step 1 2 3 4 5 6 7 8 9 10 11 Percentage 0 10 20 30 40 50 60 70 80 90 100 0 17 0 46 0 75 1 03 1 36 1 66 1 98 2 29 2 59 2 90 3 20 2 Inadequate parameter settings 0 17 0 29 0 45 0 61 0 80 1 02 1 26 1 59 2 02 2 53 3 19 3 Inadequate calibration 0 17 0 33 0 55 0 81 1 08 1 36 1 66 1 97 2 28 2 59 2 94 1 Normal 0 00 0 50 1 00 1 50 2 00 2 50 3 00 3 50 1 2 3 4 5 6 7 8 9 10 11 Step Density Normal Ina...

Page 99: ...rrection is an image correction process for monitors and film It is different from the contrast setting or grayscale setting This procedure is simply a visual adjustment As a result differences may occur depending on the operator performing the procedure Therefore be sure to consult with the responsible technician before performing this adjustment 2 4 To make the gamma of operation unit adjust in ...

Page 100: ... select END STUDY Transfer the SMPTE pattern image to the printer or the high definition monitor Fig 2 Fig 2 4 5 Take the SMPTE pattern image again On the QA screen compare the image displayed on the operation unit screen with the film image printed in Step 4 above or with the image on the high definition monitor Make sure there is no difference in contrast and gradation between those images Check...

Page 101: ... Make sure the gamma value for the test image has changed on the operation unit screen The gamma value is adjustable from 1 00 to 2 50 When the image on the operation unit is lighter whiter than the film image increase the value On the contrary if the image on the operation unit is darker blacker than the film image decrease the value The default gamma value is 1 60 Fig 4 5 3 When the gamma value ...

Page 102: ...t start up 2 3 Files are overwritten and so be careful when performing the procedure 2 4 As an extra precaution write down the numerical values before overwriting them 2 5 The overwritten counter becomes valid the next time that the CXDI software is started up 3 Procedure 3 1 Turn on the control computer and then start up Windows XP 3 2 Right click on My Computer on the Desktop screen of Windows X...

Page 103: ...OUNTER 0 TotalCount Total number of images obtained with imaging unit 1 COUNTER1 TotalCount Total number of images obtained with imaging unit 2 COUNTER2 TotalCount Total number of images obtained with imaging unit 3 COUNTER3 TotalCount Total number of images obtained with imaging unit 4 3 5 After overwriting the values overwrite the file ExpResult ini and save 3 6 Close all windows that are open o...

Page 104: ...eries Operation Manual 3 2 Backup should be performed immediately before handing over the product to the customer only for new installations 4 Connections 4 1 Check that all equipment is turned off 4 2 Connect the keyboard and mouse removal drive to the control computer 5 Settings 5 1 Turn on the operation unit power and then the control computer power 5 2 The CXDI software starts Press the Alt Ta...

Page 105: ...del number in the system BIOS settings However as a precaution check that Boot Priority Order 1 is actually set for the HDD model number If the setting is changed to Legacy Floppy Drives the control PC may not start properly when a floppy disk is inserted 2 4 Be sure to always format the floppy disk that you are using before performing backups 2 5 The backup procedure described here covers the cas...

Page 106: ...2 Installation 90 4 FD Back Up Off If FD Back Up is not necessary open the Service tool window select FD Back Up Off and press Start System Set Up menu Administrator Setup Service Tool ...

Page 107: ... BodyPart ini file is left in the BodyParts folder system will not be able to be started as the sensor type of the BodyPart and the settings do not match Example Connection np setting Sensor1 Table Table Sensor2 Stand Stand 3 Preparation 3 1 Connect the keyboard and the mouse to the control computer 3 2 Delete the ccrstart bat file from startup 3 3 Disconnect the imaging unit from the control comp...

Page 108: ...erial Number 0 1 12 0x2009 A D Board Serial Number 0 2 31 0x123 A D Board Serial Number 0 3 40G 0x123 A D Board Serial Number for SensorID 1 0x199 409 Custom Type 0 NO CUSTOM 1 STAND 2 TABLE 3 UNIV 4 CASSETTE 100um 5 CASSETTE 14X17_160um 6 CASSETTE 9X11_160um 0 0x0 0 Field of View Rotation 0 No 1 Yes 0 0x0 0 Constant for Exposure Index 0 000000 0 000000 Capture Device Configuration No 1 SensorID 2...

Page 109: ...PE BodyPart will be created appears Click OK differs according to the type of the imaging unit Æ The above message appears when a np mode is used with a setting different to that of the connected imaging unit b The messages Conflicting X ray parameters Do you wish to reset parameters and Conflicting X ray tube parameters for each imaging method Do you wish to reset parameters appear Click OK for e...

Page 110: ...e Item Operation Conditions 1 Adjust the distance between the focal point of the X ray tube and the CXDI imaging unit X cm Differs according to the grid being used 2 Perform calibration SYSTEM CALIBRATION 3 Perform self test SYSTEM SELF TEST Must pass all the tests X ray tube focal point X cm Right angle at the center CXDI Sensor Unit Self test Positioning Position the X ray tube and CXDI Imaging ...

Page 111: ...Output the image to a high resolution monitor or DICOM printer 6 Set the high resolution monitor and DICOM printer by referring to their instruction manuals 7 Check that there is no artifacts shading grid stripes etc on the image High resolution monitor DICOM printer Check image quality Exposure Out the image ...

Page 112: ...condition table msec or mAs Checking the annotation Check that the setting have been made in accordance with the customer s request IP address Subnet mask Network connections Default gateway Perform calibration Preparations prior to exposure Perform self test Checking image transfer to printers and storages Check that the setting have been made in accordance with the customer s request Checking im...

Page 113: ...Create the backup files in floppy drive by re starting Cannot be used in automobile Backing up ccr folder D ccr Check that the CXDI software starts Registering in startup Check by rebooting Check that no d np or other flags have been raised Communication with X ray generators kV mA msec body part settings etc Table 1 ...

Page 114: ...nstallation Manual 98 8 Dimension 1 Imaging Unit Mass 60G Imaging Unit 2 5 Kg w o cable 2 8 Kg inc 1 5m cable 60C Imaging Unit 2 5 Kg w o cable 2 8 Kg inc 1 5m cable Unit mm Dimension tolerance 1 Scale 1 N ...

Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...

Page 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...

Page 117: ...2 Installation Manual 101 4 Grid Optional Mass 0 6 Kg Frame only 1 1 Kg 4 1 1 3 Kg 8 1 1 4 Kg 10 1 Unit mm Dimension tolerance 1 Scale 1 N ...

Page 118: ...t C C C Ca an no on n I In nc c M Me ed di ic ca al l T Te ec ch hn ni ic ca al l S Se er rv vi ic ce e D De ep pt t A Al ll l r ri ig gh ht ts s R Re es se er rv ve ed d Printed by Canon Inc Jul 2009 Rev 01 C CX XD DI I 6 60 0G G 6 60 0C C 3 3 F Fu un nc ct ti io on n ...

Page 119: ...de 2 1 2 3 Sensor cable P630 power box side 2 1 2 4 Grid unit optional 2 1 2 5 PWB 60Di board 3 1 2 6 PCB AD REF board 3 1 2 7 PWB 60 LED board 3 2 CXDI SYSTEM II 4 2 1 POWER BOX 4 2 2 PWB 60XRAY BOARD 5 2 3 X RAY INTERFACE CABLE 5 2 4 REMOTE SWITCH 5 2 5 60 SENSOR POWER SUPPLY 5 3 ABOUT THE POWER BOX INITIALIZATION SWITCH 6 4 HOW TO ACCESS THE OS 7 ...

Page 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...

Page 121: ...g unit and Control computer via the power box Power supply from the Power box to the Imaging unit Communication between the Imaging units and Power Box The cable can be disconnected at the installation site by opening the access cover on the rear panel of the imaging unit However this job should be performed only by a service engineer or a medical engineer ME who has completed the service training...

Page 122: ...og signals that correspond to the X ray images output from the pixels on the sensor and the analog circuits which generate the reference voltages Its main functions are described below 1 Relaying the drive power to the amplifier IC 2 Relaying the control signals to the amplifier IC 3 Converting the analog signals from the amplifier IC into digital data 1 2 7 PWB 60 LED board Connected to this boar...

Page 123: ... transfer to and from the X ray generator One imaging unit can be connected to this board It is not possible to connect a multiple number of imaging units The sensor can be replaced with the other one which supports the power box by connecting the connectors of the sensor cable S150 60 imaging unit side and sensor cable P630 power box side Before the sensor is replaced the power of the power box m...

Page 124: ...sed to connect the X ray generator and power box The cable is 20 meters long Its main functions are described below Transferring the requests for X ray exposure Transferring the exposure permission to the X ray generator 2 4 Remote switch This external switch is for turning on and off the secondary output of the power box No power is supplied to the imaging unit while this switch is set to OFF but...

Page 125: ...e the initialization code when the normal code cannot start for some reason or when you have lost the Ethernet settings Using the initialization code sets the Ethernet related settings to the default settings and allows you to perform the startup operation for the initialization code In this case the connection can be made by setting the Ethernet settings in the control computer to the imaging uni...

Page 126: ...ly if the operation is not performed properly 1 3 Procedures 1 Turn ON the power of the control PC 2 When the operation screen appears press Alt and Tab keys on the keyboard together 3 Keep Alt key pressed even after the display as shown below appears Press Tab key while pressing Alt key to select the command prompt window 4 CCR Console Menu will appear Press Esc key Welcome to CCR will appear Ent...

Page 127: ...C CX XD DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 4 4 R Re ep pa ai ir r G Gu ui id de e Printed by Canon Jul 2009 Rev 01 ...

Page 128: ...able 4 2 4 Removing the PWB 60XRAY 7 3 Adhering the Grid 8 3 1 Introduction 8 3 2 Required specifications 8 3 3 Configuration 9 3 3 1 Grid unit configuration 9 3 3 2 Tools 10 3 4 Adhering the grid 11 3 4 1 Precautions 11 3 4 2 Flow of grid adhesion steps 11 3 4 3 Grid adhesion procedure 11 4 Grid Lock Assembly 18 4 1 Grid lock assembly procedure 18 4 2 Checking the grid unit installation 23 4 3 Ch...

Page 129: ...is discharged Similarly before touching the PCBs when removing them or cable connectors ensure that all static is discharged 1 3 Durability of detachable cable connector Cycle life of the detachable cable connector 4000 cycles Make sure that the power is off before connecting and disconnecting the detachable cable Replacing the Sensor Cable Periodically Communication error may occur if the connect...

Page 130: ...cover 1 Remove the six mounting screws M3x6mm for the access cover 2 Lift the access cover to remove it Notes 1 The access cover is an exterior part Handle it carefully to avoid denting or marking it 2 Be careful not to dislodge the rubber seal Screws x6 Access cover ...

Page 131: ...Attach the six mounting screws M3x6mm for the access cover Notes 1 No foreign matter must be allowed to enter 2 Be careful not to dislodge the rubber seal 3 Take care not to pinch or sandwich the rubber seal wiring or cable protection sheet 4 Check to see that the sensor cable holder is not pinched sandwiched or unsecured and that there are no gaps Rubber seal Cable protection sheet Wiring The sen...

Page 132: ...to avoid damaging the connector After disconnecting the connector check that the IC soldered to the LED board is not damaged 2 Disconnect the connector of the sensor cable which is connected to the J1 connector on the LED board Do not pull the wires to disconnect the connector Using needle nosed pliers push the three positions uniformly indicated by the red arrows in the picture below Be careful w...

Page 133: ...ect the connectors and replacement sensor cable in the sequence of steps 1 to 3 After connecting the connectors check that the IC soldered to the LED board are not damaged 5 Route the P2 connector wire bundle of the sensor cable inside the cable protection sheet Remove these two screws Cable protection sheet ...

Page 134: ...atch the edge faces of the copper foil and the cable clamp plate Fasten them by the two screws M3x6mm Fit the position of the Terminal No 1 of the P1 connector V shaped mark to the position as illustrated below Fix the sensor cable Be aware not to pinch or sandwich the wires Reference value for the tightening torque 12kgf cm 117 6N cm 4 Fasten the cable clamp upper to the cable clamp plate by the ...

Page 135: ...sconnect the cable connectors from the PWB 60XRAY 1 Distribution the two cables 2 X ray interface cable 3 Remote switch cable 4 60 power cable 2 Loosen the screw M3x6mm used to secure the PWB 60XRAY and remove it Notes 1 Do not pull the wires to disconnect the connectors 2 Be absolutely sure to disconnect the AC power cable first ...

Page 136: ...ment spacer thickness of adhesive tape If this thickness dimension of 3 5 to 3 7 mm is not satisfied unsteadiness locking difficulties or other trouble may arise when the grid unit is mounted on the imaging unit Reference CXDI 60G 60C optional grid unit The grids made by Mitaya Manufacturing Co Grid Spacer Grid ratio Focal length Thickness Spacer L Spacer S Thickness 4 1 110 cm 1 2 mm BA5 0322 BA5...

Page 137: ...liant with JIS4910 2000 3 3 Configuration 3 3 1 Grid unit configuration The parts required for adhering the grid to the grid frame unit are shown in the table below NO PARTS NAME PARTS NO Q TY REMARKS 1 Grid 1 Obtained by the sales company 2 Grid frame unit BM7 0061 1 Optional product 3 Grid spacer 4L 1 5 mm thick BA5 0322 2 4 1 grid ratio specification stamp mark 4 4 Grid spacer 4S 1 5 mm thick B...

Page 138: ...ned from one of our recommended manufacturers NO TOOL NAME TOOL NO Q TY REMARKS 1 STICKING TOOL GRID BY9 6583 1 2 WEIGHT TOOL GRID BY9 6585 1 3 Alcohol 1 For removing oil or grease on the adhesion surface 4 Scissors 1 For cutting the double sided tape Tools indicated in blue are service tools ...

Page 139: ... sticking tool on the grid frame 5 Adhere the grid to the grid frame 6 Remove the grid sticking tool from the grid frame 7 Apply pressure to the adhered grid place the grid weight tool on the grid and leave standing for 20 minutes 8 Adhere the accompanying grid labels 3 4 3 Grid adhesion procedure 1 Wipe both sides of the grid spacers short and long with ethanol to remove any oil or grease 2 Peel ...

Page 140: ...rt with the inner side of the grid sticking tool and adhere it The grid spacer must not protrude beyond the grid edge 5 Rotate the grid by 180 degrees place the tool as in step 3 and adhere the grid spacer short Grid spacer short Grid spacer short ...

Page 141: ...f the grid weight tool and adhere it Ensure that the gaps at the left and right are uniform The grid spacer must not protrude beyond the grid edge 8 Rotate the grid by 180 degrees place the tool as in step 7 provide uniform gaps at the left and right and adhere the grid spacer long Ensure that the gaps at the left and right are uniform Grid spacer long Ensure that the gaps at the left and right ar...

Page 142: ...off the backing paper on the double sided tape which has been adhered to the grid spacers Check that the double sided tape has been adhered with no parts of it standing up or wrinkled 13 Wipe around the grid frame opening with ethanol to remove any oil or grease 14 Place the grid sticking tool so that it is pressed up against the inside bending surface bottom side and left side of the grid frame D...

Page 143: ...p ends of the grid and hold the grid The grid must be installed with the serial number stamp mark at the bottom right 16 While holding the grid lower it slowly using the bottom side of the grid weight tool as the axis without applying pressure The grid must be installed with the serial number stamp mark at the bottom right Grasp both top ends of the grid Serial number stamp mark ...

Page 144: ...rid is mounted vis a vis the rows of sensor pixels However an angle of less than 0 3 degree is preferred for optimal functioning of the grid stripe reduction process 18 After applying pressure to the adhered area by hand place the grid weight tool 5 kg slowly on the grid and leave it standing for 20 minutes Take care not to allow the grid weight tool to drop onto the grid No guarantees are made fo...

Page 145: ...e grid frame where the accompanying grid labels will be adhered 20 Attach the accompanying grid labels grid specifications label and permit number label to the grid frame Reference Grid made by Mitaya Manufacturing Co 8 mm from end of inside frame 6 5 mm from end of inside frame ...

Page 146: ...ion of the anchoring hook 2 and secure the hook to the grid frame 1 using the two screws XA1 1260 406 Ensure that the anchoring hook 2 is not installed the wrong way round The anchoring hook 2 must move smoothly after the hook slide plate 3 is secured Be careful when tightening up the two screws so as to avoid damaging their heads 4 3 2 Protuberance on anchoring hook 1 ...

Page 147: ...Guide 19 2 Install the slide switch 5 on the grid frame 1 3 Temporarily secure the lock base 6 to the side switch 5 through the grid frame 1 using the screw XA4 1200 609 5 6 Temporarily secure using the screw 5 ...

Page 148: ...en up the screw XA4 1200 609 which was temporarily secured in step 3 Bear in mind that using excessive torque for tightening the screws will damage the slide switch 6 Be careful when tightening up the two screws so as to avoid damaging their heads Reference value for the tightening torque 1kgf cm 9 8N cm Press the protuberance against this part Fasten together securely using the screw 5 6 ...

Page 149: ... the lock base 6 6 Fasten the lock cover 8 securely to the grid frame 1 using the two screws XA1 1260 606 Bear in mind that using excessive torque for tightening the screws will damage the lock cover 8 Be careful when tightening up the two screws so as to avoid damaging their heads 8 6 7 ...

Page 150: ... the slide switch 5 in the unlocking direction and check that it moves smoothly together with the anchoring hook 2 to which it is linked Movement stroke of slide switch 6 mm Movement stroke of anchoring hook 4 mm 7 8 Engage the end of the spring with the hook Slide direction for unlocking Slide direction for unlocking ...

Page 151: ...t and right sides of the grid unit and fit the two tabs see detail A of the grid frame into the two cavities see detail B on the back bottom side of the imaging unit Detail C Detail D Detail B Imaging unit Cavity on reverse side Cavity on handle edge side Detail A View seen from reverse side of grid unit Detail A Grid frame tab Slide direction when unlocking Hook Detail C ...

Page 152: ... the imaging unit with the installed grid unit on a flat surface and slide the two slide switches on the outer side of the lock mechanism toward the outside This will cause the hooks fitted into the cavities of the imaging unit to be pulled up in tandem and the lock to be released 2 With the slide switches still slid hold both sides of the grid frame and separate it from the handle side of the ima...

Page 153: ...C CX XD DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 5 5 P Pa ar rt ts s C Ca at ta al lo og g Printed by Canon Jul 2009 Rev 01 ...

Page 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...

Page 155: ...ESCRIPTION MEMO 1 1 IMAGING UNIT 2 BM7 0209 000 1 CABLE UNIT SENSOR SENSOR SIDE S150 60 3 BM7 0123 000 1 CABLE UNIT SENSOR POWER BOX SIDE P630 4 BH6 5728 040 1 CABLE UNIT X RAY I F 5 BM7 0052 000 1 SWITCH UNIT REMOTE CANON LOGO BM7 0083 000 1 SWITCH UNIT REMOTE NON LOGO 3 4 5 2 1 3 ...

Page 156: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc IMAGING UNIT CXDI 60G 60C KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 0269 000 1 COVER ACCESS 2 XB1 2300 606 6 SCREW 3 XB1 2300 606 2 SCREW 1 2 3 ...

Page 157: ...ER BOX CXDI 60G 60C KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 0310 000 1 UPPER COVER POWER BOX 2 XB1 2300 406 8 SCREW 3 XB1 2300 406 4 SCREW 4 XB2 7300 606 5 SCREW W WASHER 5 BA4 1802 000 1 COVER CABLE 6 XB1 2300 406 2 SCREW 1 2 3 4 5 6 ...

Page 158: ...on Inc POWER BOX CXDI 60G 60C KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BM7 0045 000 1 PCB UNIT X RAY I F 2 BH7 9051 000 1 POWER SUPPLY UNIT 3 BM7 0046 000 1 CABLE UNIT POWER SUPPLY 4 BH7 9047 000 1 CABLE UNIT DISTRIBUTION 3 4 1 2 ...

Page 159: ...ARTS NO Q TY DESCRIPTION MEMO 1 BA5 0316 000 2 SLIDE LOCK 2 BA5 0318 000 2 COVER LOCK 3 BA4 1820 030 2 HOOK GRID 4 BA4 1822 020 2 SPRING SLIDE 5 XA1 1260 406 4 SCREW 6 XA1 1260 406 4 SCREW 7 BA4 1819 030 2 BASE HOOK SLIDE 8 BA4 1821 020 2 SPRING LOCK 2 5 6 7 1 4 3 8 ...

Page 160: ...SCREW 2 BA5 0317 000 2 BASE LOCK 3 BA5 0322 000 2 SPACER LONG t 1 5mm GRID 4 1 BA5 0325 000 2 SPACER LONG t 0 7mm GRID 8 1 BA5 0328 000 2 SPACER LONG t 0 3mm GRID 10 1 4 BA5 0323 000 2 SPACER SHORT t 1 5mm GRID 4 1 BA5 0326 000 2 SPACER SHORT t 0 7mm GRID 8 1 BA5 0329 000 2 SPACER SHORT t 0 3mm GRID 10 1 1 2 3 4 ...

Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...

Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...

Page 163: ... CX XD DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 6 6 T Tr ro ou ub bl le es sh ho oo ot ti in ng g Printed by Canon Jul 2009 Rev 01 ...

Page 164: ...curring during calibration 4 3 3 Steps to deal with errors occurring prior to ready tranfer during 5 4 Secondary Level Response 7 4 1 Steps to deal with errors occuring at host startup 7 4 2 Steps to deal with errors occurring during calibration 7 4 3 Steps to deal with errors occurring prior to ready transfer 7 4 4 Examples of abnormal images 11 5 Gathering Information About the Problem 12 6 How ...

Page 165: ...nd replaced as needed If the problem is found to be caused by the imaging unit collect the necessary data and return to the imaging unit to the distributor s office 1 2 Secondary Level Response The secondary level response is defined as the work trouble recurrence verification and repairs which is carried out by service technicians at the office of the distributor agent After it is confirmed that ...

Page 166: ... Imaging unit Collected data Action taken Primary Level Response at Hospital Service Technician ME Recurrence Verification Inspection Repair Information Acquisition date etc Secondary Level Response at distributor s office Recurrence Verification Inspection Repair Information Acquisition date etc Tertiary level response at Canon Inc Recurrence Verification Inspection Repair ...

Page 167: ...ed for the switching hub for connecting a multiple number of imaging units Category 5 or above 7 Re connect the sensor cable connections at imaging unit and power box sides 8 Check the connection of the sensor cable connections at power box side 9 Replace the sensor cable at power box side 10 Conduct a ping test check IP 192 168 100 1 11 Replace the power box 12 Replace the sensor cable at imaging...

Page 168: ... must be replaced only by a service technician who has completed the service training Steps 7 and 8 must be performed by the service technician who has been dispatched to the customer s site If the error recurs after the steps outlined above have been taken the imaging unit must be returned to the office of the distributor agent The secondary level response steps are then undertaken When a seconda...

Page 169: ... performed by the service technician who has been dispatched to the customer s site If the error recurs after the steps outlined above have been taken the imaging unit must be returned to the office of the distributor agent The secondary level response steps are then undertaken When a secondary level response is requested the following information must be provided without fail Details of how the s...

Page 170: ...g unit side must be replaced only by a service technician who has completed the service training Steps 6 and 7 must be performed by the service technician who has been dispatched to the customer s site If the abnormal images recur after the steps outlined above have been taken the imaging unit must be returned to the office of the distributor agent The secondary level response steps are then under...

Page 171: ...d power box sides connections 6 Replace the sensor cable at imaging unit side Do this if it was not done at the primary level response stage 7 Start up the host and check whether the error recurs If the error recurs after the steps outlined above have been taken the imaging unit must be returned to Canon Inc The tertiary level response steps are then undertaken When a tertiary level response is re...

Page 172: ... details on the inquiry sheet and send off the sheet Data D CCR acquired at the secondary level response stage 4 3 Steps to deal with errors occurring prior to ready transfer during ready transfer during X ray exposure and during exposure standby 1 Check for the existence of the data which was acquired at the primary level response stage Data must be acquired if it is insufficient in any way 2 Con...

Page 173: ... the sheet Data D CCR acquired at the secondary level response stage Concerning error problems Error problems may also include problems in the control computer so when an error has occurred it must be checked on the error code list of the CXDI software A tertiary level response may be requested only when it is clear that the cause of the error can be traced to the imaging unit As a basic rule requ...

Page 174: ...ensor cable at imaging unit side Do this if it was not done at the primary level response stage If the error recurs after the steps outlined above have been taken the imaging unit must be returned to Canon Inc The tertiary level response steps are then undertaken However as a basic rule requests for a tertiary level response will not be accepted for an imaging unit if the error concerned did not r...

Page 175: ...fficult to take remedial action on the market for the abnormal images shown below The problems involved preclude the possibility of repairs even when the sensor cable power box or other parts are replaced 1 Continuous multiple line defects 2 Artifacts in a block ...

Page 176: ...ure and condition setting failure during installation Electrical and mechanical system failure malfunctioning bad image noise and communication error Software and specifications problem Required Information Required log file Execute the log collection tool to save the following files 1 Environmental information such as version of the CXDI software composition of hardware optional software and the ...

Page 177: ...e system B Backup method using a hard drive The control PC has no feature that mirrors all the files including the OS to other hard drives Therefore to be ready for hard drive crashes we recommend that you add another hard drive when installing the system in order to copy the software between the hard drives using Ghost or Drive Copy which are available on the market This should be conducted at th...

Page 178: ...and then newly install the CXDI software Note Note The same version of the software must be installed 3 Copy the following files in Ccr that were copied in Step 1 to d ccr dtstore captured images Logs White dcm defpix dat xxxxxx dp 4 Newly install the optional software DMW Gen communication if you are using it 5 Copy to overwrite all the ini files in d Ccr stored on the floppy disk in order to rec...

Page 179: ...XDI software version is different When the hard drive has crashed mechanically proceed to Step 2 and subsequent steps 2 After replacement copy to overwrite all the ini files stored on the floppy disk to d ccr in order to recover the user s environment Note that images captured by users cannot be inherited in this case Note Calibration is required when the system has been used for over one year Not...

Page 180: ...e a hard drive to a different machine Required Due to being regarded as anything other than the hard drive being modified Changing external devices Not required Due to recognizing the PC s internal configuration only Adding hardware components Not required Only replacement of components that existed at the time of first activation is detected Adding components is a different category 2 Hardware co...

Page 181: ...DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 7 7 S Se er rv vi ic ce e M Ma an nu ua al l R Re ep po or rt t Printed by Canon Jul 2009 Rev 01 ...

Page 182: ...C CX XD DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 8 8 T To oo ol ls s Printed by Canon Jul 2009 Rev 01 ...

Page 183: ...DEL CANON DIGITAL RADIOGRAPHY CXDI 60G 60C BINDER SERVIC MANUAL FOR CXDI 60G 60C FILE NO 1 1 TOOL NO TOOL NAME Q TY REMARKS STICKING TOOL GRID BY9 6583 000 For CXDI 60G 60C 1 If necessary WEIGHT TOOL GRID BY9 6585 000 For CXDI 60G 60C 1 If necessary ...

Page 184: ...C CX XD DI I 6 60 0G G 6 60 0C C Canon Inc Japan Copyright C Canon Inc Medical Technical Service Dept All rights reserved 9 9 A Ap pp pe en nd di ix x Printed by Canon Jul 2009 Rev 01 ...

Page 185: ... unit 8 Imaging unit IP address Visual check 1 Connector The connector must be connected securely Visual check touch 2 Power switch It must be possible to set this switch to its ON and OFF positions properly Visual check touch 3 POWER LED The LED must light when the power is on Visual check The LED must not light when the power is on 2 Power box 4 ERROR LED The LED must light when the power is on ...

Page 186: ...no artifacts shading grid stripes etc Visual check 9 Transmission test 1 Dcap 200 times Transmission must be successful Dcap exe 10 Ping test 1 Ping 100 000 times Communication must be successful Pingtest exe 1 Dirt dust inside PC There must be no dirt or dust Visual check 2 Event logs Application logs Watson logs There must be no errors excluding errors which have been dealt with Visual check 3 S...

Reviews: