4. Repair Guide
19
4. Attachment of parts
(1) Attach aluminum spacer to the chassis of the sensor unit and then mount PCB-50LED.
Tighten the screw (XB1-2260-606) slightly weaker than usual. Otherwise, the screw hole
on the chassis of the CXDI-50G will be damaged. It is recommended to use the torque
screwdriver for tightening the screw. (Tightening torque: 4kgf•cm)
Note:
• Note that torque differs from that for the conduction tape.
•
After screws have been secured, apply adhesive
.
XD1-3100-267
XD1-2100-267
XB1-2260-606
BG7-2767-100
BA4-2434-000
Aluminum spacer
(Bent here)
Rib
Do not let the bent part of
aluminum spacer touch the rib.
PCB-50LED
Tightening torque; 4Kgf-cm
Apply adhesive.
Aluminum spacer
(Bent here)
PCB-50LED
A
A
A-A
Rib