2. Installation Manual
- 89 -
2) Look in the X-ray tube to check the cross line reflected from the mirror matches the cross line
on the X-ray tube. If it does not match, move it right and left or put the spacer before and
behind the stand base to align the slight differences.
(4) Regardless the height of the X-ray tube and the sensor unit, it should keep the
positional relation aligned in step (3) above.
x
Should be perpendicular in
any height
Center of X-ray tube and the sensor unit
should match in any height.
Case the cross line
Reflect the cross line
Correct Alignment
Shifted Horizontally
Shifted Vertically
Reflected cross line
Cross line of X-ray tube
Look in