2.1
Removing Grip unit
Base cover
Handle cap hole
Handle plate (x2)
Screw (x2)
Handle cap (x2)
Hook
(1) Remove the handle cap with the handle cap hook in the handle cap hole pushing.
(2) Loosen the screw (x2) to remove the handle plate and handle part.
Note:
1. Handle with carefully so that the defect should not adhere.
2. Check the direction of the handle cap hole at the assembling. Adjust surface which opens
the hole to the Base cover side and fix.
3. The handle cap must be pushed tightly until the hook is in the handle cap hole position.
[Replacing Handle Unit]
To improve the performance of the handle unit, following parts have been changed.
Key No.
Description
Former Part No.
New Part No.
Qty
Remarks
1
HANDLE,
IMAGING UNIT
BA4-1792-030 BA4-1792-040 1
2 CAP,
HANDLE
BA4-1793-020
BA4-1793-030
2
3
PLATE,
HANDLE
BA4-1794-020 BA4-2992-000 2
Former and new parts are not interchangeable.
Replace three parts from Key No.1 to 3 in sets to allow the compatibility with a sensor.
7M
Body No. 150827 -
CXDI-50G
3M
Body No. 250001 -
7M
Body No. 100056 -
CXDI-50C
3M
Body No. 200001 -
Summary of Contents for CXDI-50C
Page 15: ......
Page 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Page 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Page 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Page 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Page 118: ......
Page 163: ......
Page 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Page 186: ......
Page 188: ......
Page 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...