2. Installation Manual
- 86 -
4-10)When CCR SETUP MENU appears, press the [Esc] key to return to Welcome to CCR.
4-11) Select the command “8 - Exit” in the “Welcome to CCR” menu to exit the CXDI
application.
4-12) After exit the CXDI application and Windows XP desktop appears, starts the
command prompt screen (Start
⇒
Program
⇒
Command Prompt). And type the
command “ccrxxxxx /np” to start the CXDI application again.
a. When the CXDI application program is starting up, the message “There is no
BodyPart for SensorID#*. ** TYPE BodyPart will be created” appears. Click [OK].
(* differs according to the type of the sensor.)
Æ
The above message appears when a /np mode is used with a setting different to that
of the connected sensor.
b. The messages “Conflicting X-ray parameters. Do you wish to reset parameters?” and
“Conflicting X-ray tube parameters for each imaging method. Do you wish to reset
parameters” appear. Click [OK] for each.
Æ
These above messages might be appeared if /np mode is used with a different
setting.
5)
Going out of /np mode
Connect the sensor unit to the control PC.
Before using the system in normal condition, perform the following steps:
5-1) If /np mode has been used with a different setting as that of the connected sensor, as
mentioned in “(3) Notes”, delete the BodyPart**.ini file made in the BodyParts folder,
and return the BodyPart**.ini file which has been moved into the BodyParts folder.
5-2) Enter the command “ccrstart.bat” on command prompt to boot the CXDI application.
Follow the procedure from 3) to 9) in previous section. And check these one will be
preformed properly without any problems.
5-3) Register the ccrstart.bat file to the StartUp.
Summary of Contents for CXDI-50C
Page 15: ......
Page 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Page 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Page 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Page 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Page 118: ......
Page 163: ......
Page 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Page 186: ......
Page 188: ......
Page 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...