3. Function
10
2. Imaging condition check at the time of upgrade
If any of the kV, mA, mAs, cm (except msec) values recorded in the imaging condition
table or in each body part information exceeds the limit of the sensor, when the CXDI
starts up, a warning message will be displayed (Fig. 1) and the value will be automatically
modified to be at or below the limit. (However, thinking from the range of values that can
be used with the generator, the probability of the warning message is very low.)
3. Adding values to the imaging conditions table
It is possible to add values up to 3000 msec, 4800 mAs to the imaging conditions table.
However, at the time of the actual imaging, if the exposure is performed with a CXDI-11 -
CXDI-31 sensor, it is not possible to set values higher than 1000 msec, 1600 mAs.
Figure 3-1