4. Repair Guide
7
2.4
Exchanging the sensor cable
(1) Removing the Cable
1.
Remove the screws (M3x8) x2 that fixes the cable clamp 1.
2.
Remove the screws (M2.6x6) x2 that fixes the cable holder.
3.
Remove the screws (M2.6x6) x2 that fixes the cable clamp 2. Do not lose the screw
because it is held by the clamp cushion.
4.
Disconnect the connectors from PCB-50LED and PWB-50Di.
5.
Remove the sensor cable unit.
NOTES:
1.
During assembly, lay out the signal cables carefully to they will not get caught by
covers.
Sensor cable
Screw M2.6x6 (x2)
Cable clamp 2
Screw M3x8 (x2)
Cable clamp 1
Disconnect the signal
cable connector.
Screw M2.6x6 (x2)
Clamp cushion
The cable protect sheet is put
under the cable so as to protect
the sensor cable.
Bush
Disconnect the power
cable connector.
Be careful not to get the cable
caught between covers.
Bend the power line.
Cable holder