2. Installation Manual
- 73 -
4-3) Select the END STUDY and transfer the SMPTE pattern image to the printer or the
high-definition monitor. [Fig. 6]
[Fig. 6]
4-4) On a printed film or on the monitor, measure the density of 32 steps of the grayscale
on the test image. [Fig. 7]
*1: Measure the SMPTE image density on the film using a densitometer. The SMPTE
image density on the high-definition monitor is measured using the gradation
analysis software.
*2: The data for the SMPTE test image grayscale transferred by the CXDI are the
values for the maximum density (3.20 in the case of the MLP 190) in the printer
settings which have been changed into 32 uniform steps.
[Fig. 7]
Step1
Step32
Summary of Contents for CXDI-50C
Page 15: ......
Page 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Page 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Page 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Page 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Page 118: ......
Page 163: ......
Page 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Page 186: ......
Page 188: ......
Page 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...