2. Installation Manual
- 93 -
5.9 Image Quality
(1) Purpose
This procedure is used to check the final image quality of the CXDI.
1) Resolution
check
Tools used
(1) Phantom
(2) High-resolution monitor or DICOM printer
Procedure
Procedure/Item Operation/Conditions
1. Adjust the distance between the focal point of the X-
ray tube and the CXDI sensor unit.
X cm: Differs according to the grid being used.
2. Perform calibration
SYSTEM > CALIBRATION
3. Perform self-test.
SYSTEM > SELF TEST
Must pass all the tests.
X-ray tube focal
point
X cm
Right angle at
the center
CXDI Sensor Unit
Self-test
[Positioning]
Position the X-ray tube and
CXDI Sensor Unit
Calibration